Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1120

Current through Register Vol. 49, No. 29, 2025
Part 4732.1120 - THERAPEUTIC EQUIPMENT PERFORMANCE TESTS AND LIMITS FOR MEASUREMENT EQUIPMENT
Subpart 1.Required tests.
A. Installation calibration tests must be conducted prior to any patient use. Any adjustments must be made to bring the equipment up to a nationally recognized standard, such as Code of Federal Regulations, title 21, section 892, or manufacturer's specifications and to ensure compliance with this chapter prior to first use.
B. Equipment performance tests must be conducted over all clinical ranges, when applicable. For equipment performance tests, any adjustments must be made to bring equipment to compliance with a nationally recognized standard, such as Code of Federal Regulations, title 21, section 892, or manufacturer's specifications and to ensure compliance with this chapter prior to using the equipment again.
Subp. 2.Local standard (Loc Std) instrument.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. AAPM - accredited dosimetry calibration laboratory calibrationIntercomparison every 12 months At intervals not to exceed 24 months traceable to NIST StandardDocumented and correction applied or noted in report of measurement when appropriate
B. LinearityAt intervals not to exceed 48 months0.5 percent
C. VentingAt intervals not to exceed 48 monthsDocumented and correction applied
D. Extracameral signalInitial use0.5 percent
E. LeakageEach use0.1 percent
F. RecombinationInitial useDocumented and correction applied
G. Collecting potentialEach useDocumented and correction applied

Subp. 3.Other field instruments.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Local standard comparisonAt intervals not to exceed 24 months1 percent
B. LinearityAt intervals not to exceed 24 monthsDocumented and correction applied
C. VentingAt intervals not to exceed 24 monthsDocumented and correction applied
D. Extracameral signalAt intervals not to exceed 24 monthsDocumented and correction applied
E. LeakageEach use1 percent
F. RecombinationInitial useDocumented and correction applied
G. Collecting potentialEach useDocumented and correction applied

Subp. 4.Relative dosimetry equipment.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Thermoluminescent Dosimeter
(1) CalibrationEach batch or boxDocumented and correction applied
(2) LinearityInitial useDocumented and correction applied
B. Film
(1) Dose and responseEach batch or boxDocumented and correction applied
(2) Densitometer linearityAt intervals not to exceed 12 monthsDocumented and correction applied
C. Air ionization chamber system
(1) LinearityAt intervals not to exceed 12 monthsDocumented and correction applied
(2) Extracameral signalInitial use1 percent
D. Diode system
(1) Energy dependenceInitial useDocumented and correction applied
(2) Extracameral signalInitial useDocumented and correction applied
(3) LinearityInitial useDocumented and correction applied

Subp. 5.Radiation survey instruments.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. CalibrationAt intervals not to exceed 12 monthsDocumented and correction applied
B. LinearityAt intervals not to exceed 12 monthsDocumented and correction applied
C. ConstancyEach use5 percent

Subp. 6.Positioning equipment lasers.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. AccuracyDaily before patient use2 mm
B. HysteresisEach use2 mm

Subp. 7.Phantoms and attenuators.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. ThicknessInitial useDocumented and correction applied
B. DensityInitial useDocumented and correction applied
C. Phantom stacked densityInitial useDocumented and correction applied
D. Detector fitInitial useDocumented and correction applied

Subp. 8.Accessory equipment.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Thermometer calibrationInitial use0.1 degree/C
B. Barometer (aneroid)
(1) Calibration HgInitial use1 mm Hg
(2) IntercomparisonAt intervals not to exceed 12 months1 mm Hg

Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1120

32 SR 777

Statutory Authority: MS s 144.12