TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. AAPM - accredited dosimetry calibration laboratory calibration | Intercomparison every 12 months At intervals not to exceed 24 months traceable to NIST Standard | Documented and correction applied or noted in report of measurement when appropriate |
B. Linearity | At intervals not to exceed 48 months | 0.5 percent |
C. Venting | At intervals not to exceed 48 months | Documented and correction applied |
D. Extracameral signal | Initial use | 0.5 percent |
E. Leakage | Each use | 0.1 percent |
F. Recombination | Initial use | Documented and correction applied |
G. Collecting potential | Each use | Documented and correction applied |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Local standard comparison | At intervals not to exceed 24 months | 1 percent |
B. Linearity | At intervals not to exceed 24 months | Documented and correction applied |
C. Venting | At intervals not to exceed 24 months | Documented and correction applied |
D. Extracameral signal | At intervals not to exceed 24 months | Documented and correction applied |
E. Leakage | Each use | 1 percent |
F. Recombination | Initial use | Documented and correction applied |
G. Collecting potential | Each use | Documented and correction applied |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Thermoluminescent Dosimeter | ||
(1) Calibration | Each batch or box | Documented and correction applied |
(2) Linearity | Initial use | Documented and correction applied |
B. Film | ||
(1) Dose and response | Each batch or box | Documented and correction applied |
(2) Densitometer linearity | At intervals not to exceed 12 months | Documented and correction applied |
C. Air ionization chamber system | ||
(1) Linearity | At intervals not to exceed 12 months | Documented and correction applied |
(2) Extracameral signal | Initial use | 1 percent |
D. Diode system | ||
(1) Energy dependence | Initial use | Documented and correction applied |
(2) Extracameral signal | Initial use | Documented and correction applied |
(3) Linearity | Initial use | Documented and correction applied |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Calibration | At intervals not to exceed 12 months | Documented and correction applied |
B. Linearity | At intervals not to exceed 12 months | Documented and correction applied |
C. Constancy | Each use | 5 percent |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Accuracy | Daily before patient use | 2 mm |
B. Hysteresis | Each use | 2 mm |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Thickness | Initial use | Documented and correction applied |
B. Density | Initial use | Documented and correction applied |
C. Phantom stacked density | Initial use | Documented and correction applied |
D. Detector fit | Initial use | Documented and correction applied |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Thermometer calibration | Initial use | 0.1 degree/C |
B. Barometer (aneroid) | ||
(1) Calibration Hg | Initial use | 1 mm Hg |
(2) Intercomparison | At intervals not to exceed 12 months | 1 mm Hg |
Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1120