Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1130

Current through Register Vol. 49, No. 24, December 9, 2024
Part 4732.1130 - EQUIPMENT PERFORMANCE TESTS FOR EXTERNAL BEAM TELETHERAPY AND SIMULATION SYSTEMS
Subpart 1.Dosimetry.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Central axis dose calibrationAt intervals not to exceed 12 months2 percent
B. Constancy checks-photons
(1) Dose per monitor unit along central axisWeekly3 percent
(2) Depth doseMonthly2 percent
(3) Beam uniformityMonthly3 percent
(4) Monitor chamber linearityAt intervals not to exceed 12 months1 percent
(5) Timer linearity and errorAt intervals not to exceed 12 months1 percent

Subp. 2.Geometry.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Field positioning aids
(1) Light field and radiation field agreementMonthly2 mm
(2) Mechanical distance pins, lasers, and SSD lightsMonthly2 mm
(3) Scale readoutsMonthly2 mm/1 degree angle
B. Machine alignment
(1) Jaw symmetryAt intervals not to exceed 12 months2 mm
(2) Coincidence of collimator (jaw) and gantry axesAt intervals not to exceed 12 months2 mm
(3) Stability of gantry arm and bearing under rotationAt intervals not to exceed 12 months2 mm
(4) Couch motion and tabletop sagAt intervals not to exceed 12 months2 mm

Subp. 3.Constancy checks-electrons.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Beam uniformityMonthly5 percent
B. Depth doseMonthly2 mm at therapeutic depth
C. Dose per monitor unit constancy checkWeekly3 percent

Subp. 4.Treatment accessories*.

TEST TYPEMINIMUM TEST INTERVALTOLERANCE
A. Wedge transmission factorAt intervals not to exceed 12 months2 percent
B. Transmission factor constancy for all treatment accessoriesAt intervals not to exceed 12 months2 percent

*Attenuation in blocks, wedge factors, and compensator data must be checked annually. A visual inspection of the mechanical integrity of these accessories must be done monthly.

Subp. 5.Simulators.

TEST TYPEFREQUENCYTOLERANCE
A. Localizing lasersDaily2 mm
B. Distance indicatorDaily2 mm
C. Field size indicatorMonthly2 mm
D. Gantry/collimator angle indicatorsMonthly1 degree
E. Cross-hair centeringMonthly2 mm diameter
F. Focal spot-axis indicatorMonthly2 mm
G. Fluoroscopic image qualityMonthlyEstablished baseline
H. Collision avoidanceMonthlyFunctional
I. Light/radiation field coincidenceMonthly2 mm or 1 percent
J. Collimator rotation isocenterAt intervals not to exceed 12 months2 mm diameter
K. Gantry rotation isocenterAt intervals not to exceed 12 months2 mm diameter
L. Couch rotation isocenterAt intervals not to exceed 12 months2 mm diameter
M. Coincidence of collimator, gantry, couch axes, and isocenterAt intervals not to exceed 12 months2 mm diameter
N. Table top sagAt intervals not to exceed 12 months2 mm
O. Vertical travel of couchAt intervals not to exceed 12 months2 mm
P. Exposure rateAt intervals not to exceed 12 monthsEstablished baseline
Q. Table top exposure with fluoroscopyAt intervals not to exceed 12 monthsEstablished baseline
R. kVp and mAs calibrationAt intervals not to exceed 12 monthsEstablished baseline
S. High and low contrast resolutionAt intervals not to exceed 12 monthsEstablished baseline

Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1130

32 SR 777

Statutory Authority: MS s 144.12