The tests in this part are to be made at the time of installation and at the specified intervals thereafter.
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Screen-film contact | At intervals not to exceed 24 months | No significant areas of poor contact as measured by no less than: (1) 8 wires/inch mesh; or (2) 7 holes/inch for regular film; (3) 40 wires/inch mesh or greater for mammography film |
B. | Screen-film- cassette speed match | At intervals not to exceed 24 months | Densities within ± 0.10 O.D. for all cassettes of the same speed used for imaging |
C. | CR imaging plates | At intervals not to exceed three months or upon observation of image artifacts | Follow manufacturer's recommendations |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Darkroom fog | At intervals not to exceed six months | < 0.08 O.D. increase in density (measured at approximately 1.00 O.D.) after 2 minutes using film exposed on-site at the time of test. For mammography the O.D. increase must be < 0.05 |
B. | Sensitometry and densitometry | Before processing first film of the day | Density difference ± 0.15 O.D. and base + fog + .05 O.D. using film exposed on-site at time of test. Veterinary facilities are not required to perform this test |
C. | Temperature check | At the time of sensitometry | Follow manufacturer's recommendations |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | SID indicator accuracy | At intervals not to exceed 24 months | ± 2% of indicated value |
B. | X-ray and light field alignment | At intervals not to exceed 24 months | ± 2% of SID any one direction, ± 3% of SID, both directions (total) |
C. | X-ray and image receptor alignment | At intervals not to exceed 24 months | ± 2% of SID |
D. | Collimator dial accuracy | At intervals not to exceed 24 months | ± 2% of SID |
E. | Reproducibility | At intervals not to exceed 24 months | Coefficient of variation < 5% |
F. | mR/mAs | At intervals not to exceed 24 months | ± 10% of baseline |
G. | Linearity | At intervals not to exceed 24 months | ± 10% over clinical range |
H. | Linearity - for mAs only units manufactured after May 3, 1994 | At intervals not to exceed 24 months | Average ratios of exposure to the indicated mAs obtained in any two consecutive mAs settings must not differ by more than 0.10 times their sum, or at two settings differing by no more than a factor of two where the mAs selector provides continuous selection |
I. | Timer accuracy | At intervals not to exceed 24 months | Single Phase: ± 10% of setting. Three phase, high frequency, and constant potential: use ± 5% of selected time when measured > 100 milliseconds. At times shorter than 100 milliseconds, use manufacturers' specifications |
J. | Half-value layer | At intervals not to exceed 24 months | Must meet requirements in part 4732.0810 |
K. | kVp accuracy | At intervals not to exceed 24 months | ± 5% of indicated kVp |
L. | Phototimer reproducibility, if present | At intervals not to exceed 24 months | ± 5% of average exposure |
M. | AEC (phototimer) | At intervals not to exceed 24 months | ± 10% of manufacturer's state increments |
N. | Illuminance of collimator | At intervals not to exceed 24 months | > 15 footcandles |
O. | Film density vs. thickness change on AEC | At intervals not to exceed 24 months | ± 0.30 O.D. of the averaged exposures over the range specified by the manufacturer |
P. | Film density vs. kVp change on AEC | At intervals not to exceed 24 months | ± 0.30 O.D. of the averaged exposures when measured at > 1.2 O.D. and over the range as specified by the manufacturer |
Q. | Spot film reproducibility (fluoroscopy units with manual mode) | At intervals not to exceed 24 months | ± 5% of average exposure |
R. | Phototimer back-up timer cut off | At time of installation | Terminates exposure at < 600 mAs |
S. | AEC density at normal or "0" | At intervals not to exceed 24 months | > 1.0 O.D. |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Maximum output at tabletop or equivalent minimum SSD | At intervals not to exceed 12 months and every tube change | < 5 R (1.3 mC/kg) per minute for manual; < 10 R (2.6 mC/kg) per minute for automatic exposure rate control systems |
B. | High level control maximum output at tabletop or equivalent minimum SSD | At intervals not to exceed 12 months and every tube change | < 20 R (5.0 mC/kg-1) per minute |
C. | Fluoroscopic image size | At intervals not to exceed 12 months and every tube change | Error between fluorographic beam size and observed image size must be no more than ± 3% of SID for all modes and at any tower height |
D. | Actual spot-film size vs. indicated | At intervals not to exceed 12 months | Error between actual fluorographic beam size at image receptor and indicated image size must be no more than ± 3% of SID for all modes and at any tower height |
E. | Spot-film reproducibility | At intervals not to exceed 12 months | ± 5% of average exposure |
F. | Phototimer reproducibility, if present | At intervals not to exceed 12 months | ± 5% of average exposure |
G. | Fluoroscopic high contrast resolution and distortion | At intervals not to exceed 12 months | Six inch (15 centimeter) intensifier: center 30 and edge 24 (wires per inch) copper mesh; nine inch (23 centimeter) intensifier |
H. | Half-value layer | At intervals not to exceed 12 months and after every tube change | ± 5% for equipment manufactured before 1973. For equipment manufactured after 1973, follow manufacturer's specified limits |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Maximum output at tabletop or equivalent minimum SSD | At intervals not to exceed 12 months and at every tube change | > 5 R/min must have automatic exposure rate control; > 10 R/min must have high level control; if not high level control maximum is < 10 R/min |
B. | High level control maximum output at tabletop or equivalent minimum SSD | At intervals not to exceed 12 months and at every tube change | < 20 R/min |
C. | All other tests as indicated in subpart 5 | At intervals not to exceed 24 months | See criteria in subpart 5 |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Section level | At intervals not to exceed 12 months | ± 5 millimeters |
B. | Level incrementation | At intervals not to exceed 12 months | ± 2 millimeters |
C. | Section thickness (slice width) | At intervals not to exceed 12 months | Follow manufacturer's specifications |
D. | All other tests in part 4732.1000 if applicable | At intervals not to exceed 24 months | See criteria in subpart 4 |
E. | Spatial plane resolution | At intervals not to exceed 12 months | 40 mesh screen or better |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Accuracy of scout localization view | At intervals not to exceed 12 months | ± 1 millimeters |
B. | Accuracy of distance measurements | At intervals not to exceed 12 months | ± 1 millimeters |
C. | CT dose index | At intervals not to exceed 12 months | ± 20% from manufacturer's recommendations |
D. | CT number dependence on slice thickness | At intervals not to exceed 12 months | Mean ± 3 CT numbers averaged over 100 pixels |
E. | CT number calibration and noise | Daily | Water: 0 ± 5 CT numbers; Noise: ± 3 standard deviations of the mean of the baseline noise variance measurements |
F. | CT number uniformity and artifacts | Monthly for mobile units. At intervals not to exceed 12 months for fixed base units. | Variation ± 5 CT numbers between the mean values of measurements made at center and edge of phantom that is at least 20 cm. In diameter among a mean of 100 pixels. Artifacts: no noticeable artifacts |
G. | Hard copy output and visual display | Daily | Luminance and contrast not significantly different |
H. | Table indexing | At intervals not to exceed six months | ± 0.5 millimeter for each increment |
I. | Table backlash | At intervals not to exceed six months | ± one millimeter |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Cinefluorographic exposure rates | At intervals not to exceed 12 months | Approximately 10 to 20 µR (2.6 to 5.0 nC/kg) per frame at intensifier for nine inch (23 cm) mode; approximately 20 to 30 µR (5 to 8 nC/kg) per frame at intensifier for six inch (15 cm) mode |
B. | All tests in subparts 4, 5, and 6, if applicable | At intervals not to exceed 24 months | See criteria in subparts 4, 5, and 6 |
C. | Film changer screen-film contact | At intervals not to exceed 24 months | No significant areas of poor contact as measured by no less than: (1) 8 wire per inch mesh; or (2) 7 holes per inch |
D. | High contrast resolution for cinefluorographic and digital systems | At intervals not to exceed 12 months | No significant difference between static and dynamic conditions |
E. | Optical density of films over duration of filming run | At intervals not to exceed 12 months | < ± 0.2 O.D. difference |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Film processing | Before the first film of the day | Between 0.75 and 1.05 O.D. on the test tool or follow test tool manufacturer's recommendations |
B. | Fog test | At intervals not to exceed six months | Unable to visualize coin edges |
C. | Filtration (HVL) | At intervals not to exceed 24 months | Meet requirements in part 4732.0800 |
D. | Radiation exposure at the end of cone | At intervals not to exceed 24 months | Meet requirements in part 4732.0825 |
E. | Timer reproducibility | At intervals not to exceed 24 months | ± 10% of indicated timer setting |
F. | kVp accuracy | At intervals not to exceed 24 months | ± 5% of indicated kVp for equipment manufactured before 1973. For equipment manufactured after 1973, follow manufacturer's specified limits |
G. | Exposure output reproducibility | At intervals not to exceed 24 months | Coefficient of variation < 5% |
H. | Dental mA linearity | At intervals not to exceed 24 months | ± 10% over the clinical range |
TEST TYPE | MINIMUM TEST INTERVAL | MINIMUM PERFORMANCE CRITERIA | |
A. | Film processing | Before the first film of the day | Use processing as specified in subpart 3. A step wedge may be used. ± one step from standard allowed |
B. | Fog test | At intervals not to exceed six months | Use criteria in subpart 3, item A, for automatic processing; subpart 4, item A, for manual processing |
C. | Same test types and minimum performance criteria as in diagnostic radiographic tubes in subpart 4 | At intervals not to exceed 24 months | See criteria in subpart 4 |
Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1100
Statutory Authority: MS s 144.12