Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1100

Current through Register Vol. 49, No. 24, December 9, 2024
Part 4732.1100 - INSTALLATION CALIBRATION TESTS AND EQUIPMENT PERFORMANCE TESTS FOR A QUALITY ASSURANCE PROGRAM
Subpart 1.Tests required.
A. Installation calibration tests must be conducted prior to any patient use. Any adjustments must be made to bring the equipment up to a nationally recognized standard such as Code of Federal Regulations, title 21, section 1020, or the manufacturer's specifications, and to ensure compliance with this chapter prior to first use.
B. Equipment performance tests must be conducted over all clinical ranges, when applicable. For equipment performance tests, any adjustments must be made to bring equipment to a nationally recognized standard or manufacturer's specifications; and to ensure compliance with this chapter prior to using the equipment again.
Subp. 2.Frequency of tests.

The tests in this part are to be made at the time of installation and at the specified intervals thereafter.

Subp. 3.Image receptors.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Screen-film contactAt intervals not to exceed 24 monthsNo significant areas of poor contact as measured by no less than: (1) 8 wires/inch mesh; or (2) 7 holes/inch for regular film; (3) 40 wires/inch mesh or greater for mammography film
B.Screen-film- cassette speed matchAt intervals not to exceed 24 monthsDensities within ± 0.10 O.D. for all cassettes of the same speed used for imaging
C.CR imaging platesAt intervals not to exceed three months or upon observation of image artifactsFollow manufacturer's recommendations

Subp. 4.Processing.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Darkroom fogAt intervals not to exceed six months< 0.08 O.D. increase in density (measured at approximately 1.00 O.D.) after 2 minutes using film exposed on-site at the time of test. For mammography the O.D. increase must be < 0.05
B.Sensitometry and densitometryBefore processing first film of the dayDensity difference ± 0.15 O.D. and base + fog + .05 O.D. using film exposed on-site at time of test. Veterinary facilities are not required to perform this test
C.Temperature checkAt the time of sensitometryFollow manufacturer's recommendations

Subp. 5.All diagnostic radiographic tubes; required when applicable.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.SID indicator accuracyAt intervals not to exceed 24 months ± 2% of indicated value
B.X-ray and light field alignmentAt intervals not to exceed 24 months ± 2% of SID any one direction, ± 3% of SID, both directions (total)
C.X-ray and image receptor alignmentAt intervals not to exceed 24 months ± 2% of SID
D.Collimator dial accuracyAt intervals not to exceed 24 months ± 2% of SID
E.ReproducibilityAt intervals not to exceed 24 monthsCoefficient of variation < 5%
F.mR/mAsAt intervals not to exceed 24 months ± 10% of baseline
G.LinearityAt intervals not to exceed 24 months ± 10% over clinical range
H.Linearity - for mAs only units manufactured after May 3, 1994At intervals not to exceed 24 monthsAverage ratios of exposure to the indicated mAs obtained in any two consecutive mAs settings must not differ by more than 0.10 times their sum, or at two settings differing by no more than a factor of two where the mAs selector provides continuous selection
I.Timer accuracyAt intervals not to exceed 24 monthsSingle Phase: ± 10% of setting. Three phase, high frequency, and constant potential: use ± 5% of selected time when measured > 100 milliseconds. At times shorter than 100 milliseconds, use manufacturers' specifications
J.Half-value layerAt intervals not to exceed 24 monthsMust meet requirements in part 4732.0810
K.kVp accuracyAt intervals not to exceed 24 months ± 5% of indicated kVp
L.Phototimer reproducibility, if presentAt intervals not to exceed 24 months ± 5% of average exposure
M.AEC (phototimer)At intervals not to exceed 24 months ± 10% of manufacturer's state increments
N.Illuminance of collimatorAt intervals not to exceed 24 months> 15 footcandles
O.Film density vs. thickness change on AECAt intervals not to exceed 24 months ± 0.30 O.D. of the averaged exposures over the range specified by the manufacturer
P.Film density vs. kVp change on AECAt intervals not to exceed 24 months ± 0.30 O.D. of the averaged exposures when measured at > 1.2 O.D. and over the range as specified by the manufacturer
Q.Spot film reproducibility (fluoroscopy units with manual mode)At intervals not to exceed 24 months ± 5% of average exposure
R.Phototimer back-up timer cut offAt time of installationTerminates exposure at < 600 mAs
S.AEC density at normal or "0"At intervals not to exceed 24 months> 1.0 O.D.

Subp. 6.For facilities with fluoroscopes and C-arm fluoroscopes, except radiation therapy simulators, manufactured before May 19, 1995.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Maximum output at tabletop or equivalent minimum SSDAt intervals not to exceed 12 months and every tube change< 5 R (1.3 mC/kg) per minute for manual; < 10 R (2.6 mC/kg) per minute for automatic exposure rate control systems
B.High level control maximum output at tabletop or equivalent minimum SSDAt intervals not to exceed 12 months and every tube change< 20 R (5.0 mC/kg-1) per minute
C.Fluoroscopic image sizeAt intervals not to exceed 12 months and every tube changeError between fluorographic beam size and observed image size must be no more than ± 3% of SID for all modes and at any tower height
D.Actual spot-film size vs. indicatedAt intervals not to exceed 12 monthsError between actual fluorographic beam size at image receptor and indicated image size must be no more than ± 3% of SID for all modes and at any tower height
E.Spot-film reproducibilityAt intervals not to exceed 12 months ± 5% of average exposure
F.Phototimer reproducibility, if presentAt intervals not to exceed 12 months ± 5% of average exposure
G.Fluoroscopic high contrast resolution and distortionAt intervals not to exceed 12 monthsSix inch (15 centimeter) intensifier: center 30 and edge 24 (wires per inch) copper mesh; nine inch (23 centimeter) intensifier
H.Half-value layerAt intervals not to exceed 12 months and after every tube change ± 5% for equipment manufactured before 1973. For equipment manufactured after 1973, follow manufacturer's specified limits

Subp. 7.For facilities with fluoroscopes and C-arm fluoroscopes, except radiation therapy simulators, manufactured on or after May 19, 1995.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Maximum output at tabletop or equivalent minimum SSDAt intervals not to exceed 12 months and at every tube change> 5 R/min must have automatic exposure rate control; > 10 R/min must have high level control; if not high level control maximum is < 10 R/min
B.High level control maximum output at tabletop or equivalent minimum SSDAt intervals not to exceed 12 months and at every tube change< 20 R/min
C.All other tests as indicated in subpart 5At intervals not to exceed 24 monthsSee criteria in subpart 5

Subp. 8.For facilities with tomography systems other than computed tomography.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Section levelAt intervals not to exceed 12 months ± 5 millimeters
B.Level incrementationAt intervals not to exceed 12 months ± 2 millimeters
C.Section thickness (slice width)At intervals not to exceed 12 monthsFollow manufacturer's specifications
D.All other tests in part 4732.1000 if applicableAt intervals not to exceed 24 monthsSee criteria in subpart 4
E.Spatial plane resolutionAt intervals not to exceed 12 months40 mesh screen or better

Subp. 9.For facilities with computed tomography scanners.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Accuracy of scout localization viewAt intervals not to exceed 12 months ± 1 millimeters
B.Accuracy of distance measurementsAt intervals not to exceed 12 months ± 1 millimeters
C.CT dose indexAt intervals not to exceed 12 months ± 20% from manufacturer's recommendations
D.CT number dependence on slice thicknessAt intervals not to exceed 12 monthsMean ± 3 CT numbers averaged over 100 pixels
E.CT number calibration and noiseDailyWater: 0 ± 5 CT numbers; Noise: ± 3 standard deviations of the mean of the baseline noise variance measurements
F.CT number uniformity and artifactsMonthly for mobile units. At intervals not to exceed 12 months for fixed base units.Variation ± 5 CT numbers between the mean values of measurements made at center and edge of phantom that is at least 20 cm. In diameter among a mean of 100 pixels. Artifacts: no noticeable artifacts
G.Hard copy output and visual displayDailyLuminance and contrast not significantly different
H.Table indexingAt intervals not to exceed six months ± 0.5 millimeter for each increment
I.Table backlashAt intervals not to exceed six months ± one millimeter

Subp. 10.For facilities with cinefluorographic and special procedure systems.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Cinefluorographic exposure ratesAt intervals not to exceed 12 monthsApproximately 10 to 20 µR (2.6 to 5.0 nC/kg) per frame at intensifier for nine inch (23 cm) mode; approximately 20 to 30 µR (5 to 8 nC/kg) per frame at intensifier for six inch (15 cm) mode
B.All tests in subparts 4, 5, and 6, if applicableAt intervals not to exceed 24 monthsSee criteria in subparts 4, 5, and 6
C.Film changer screen-film contactAt intervals not to exceed 24 monthsNo significant areas of poor contact as measured by no less than: (1) 8 wire per inch mesh; or (2) 7 holes per inch
D.High contrast resolution for cinefluorographic and digital systemsAt intervals not to exceed 12 monthsNo significant difference between static and dynamic conditions
E.Optical density of films over duration of filming runAt intervals not to exceed 12 months< ± 0.2 O.D. difference

Subp. 11.For facilities with dental intraoral systems.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Film processingBefore the first film of the dayBetween 0.75 and 1.05 O.D. on the test tool or follow test tool manufacturer's recommendations
B.Fog testAt intervals not to exceed six monthsUnable to visualize coin edges
C.Filtration (HVL)At intervals not to exceed 24 monthsMeet requirements in part 4732.0800
D.Radiation exposure at the end of coneAt intervals not to exceed 24 monthsMeet requirements in part 4732.0825
E.Timer reproducibilityAt intervals not to exceed 24 months ± 10% of indicated timer setting
F.kVp accuracyAt intervals not to exceed 24 months ± 5% of indicated kVp for equipment manufactured before 1973. For equipment manufactured after 1973, follow manufacturer's specified limits
G.Exposure output reproducibilityAt intervals not to exceed 24 monthsCoefficient of variation < 5%
H.Dental mA linearityAt intervals not to exceed 24 months ± 10% over the clinical range

Subp. 12.For facilities with dental extraoral systems including panoramic systems.

TEST TYPEMINIMUM TEST INTERVALMINIMUM PERFORMANCE CRITERIA
A.Film processingBefore the first film of the dayUse processing as specified in subpart 3. A step wedge may be used. ± one step from standard allowed
B.Fog testAt intervals not to exceed six monthsUse criteria in subpart 3, item A, for automatic processing; subpart 4, item A, for manual processing
C.Same test types and minimum performance criteria as in diagnostic radiographic tubes in subpart 4At intervals not to exceed 24 monthsSee criteria in subpart 4

Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1100

32 SR 777

Statutory Authority: MS s 144.12