S.D. Admin. R. 44:03:01:12.01

Current through Register Vol. 51, page 67, December 16, 2024
Section 44:03:01:12.01 - Radiation producing equipment calibration

The licensee shall provide that calibrations are performed on a diagnostic radiographic system if that system does not meet the minimum performance criteria specified in §§ 44:03:01:08.02 to 44:03:01:08.07, inclusive, and if there is any change or replacement of components that could cause a change in the radiation output of that system.

The calibration may not exceed three months after any change or replacement of components that could cause a change in the radiation output. The calibration of the radiation output of the X ray system shall be performed by or under the direction of a qualified expert. Calibration of the radiation output of an X ray system shall be performed with a calibrated dosimetry system. Any X ray machine shall be surveyed by the department of a qualified expert every three years to verify the X ray machine meets the standards of this chapter. Any X ray machine not meeting the standards of this chapter must be calibrated. Any computed tomography system shall be surveyed by a medical physicist prior to clinical use and on an annual basis.

S.D. Admin. R. 44:03:01:12.01

26 SDR 96, effective 1/23/2000; 32 SDR 128, effective 1/30/2006; 50 SDR 041, effective 10/11/2023

General Authority: SDCL 34-1-17, 34-21-4.1, 34-21-15.

Law Implemented: SDCL 34-21-4.1, 34-21-15.