Current through Register Vol. 51, page 67, December 16, 2024
Section 44:03:01:08.03 - Equipment standards for medical fluoroscopic X ray machines The standards for fluoroscopic X ray equipment are:
(1) All medical fluoroscopic X ray machines must meet the specifications in 21 C.F.R. § 1020.30, in effect on March 31, 2000, and 21 C.F.R. § 1020.32, in effect on May 19, 1994;(2) Any exposure to the operator's eyes above the screen and to the operator's waist behind the leaded drapes may not exceed fifty milliroentgens per hour;(3) A dead-man switch shall control the fluoroscopic device.S.D. Admin. R. 44:03:01:08.03
26 SDR 96, effective 1/23/2000; 50 SDR 041, effective 10/11/2023General Authority: SDCL 34-21-4.1, 34-21-15.
Law Implemented: SDCL 34-21-4.1, 34-21-15.