S.D. Admin. R. 44:03:01:08.03

Current through Register Vol. 51, page 67, December 16, 2024
Section 44:03:01:08.03 - Equipment standards for medical fluoroscopic X ray machines

The standards for fluoroscopic X ray equipment are:

(1) All medical fluoroscopic X ray machines must meet the specifications in 21 C.F.R. § 1020.30, in effect on March 31, 2000, and 21 C.F.R. § 1020.32, in effect on May 19, 1994;
(2) Any exposure to the operator's eyes above the screen and to the operator's waist behind the leaded drapes may not exceed fifty milliroentgens per hour;
(3) A dead-man switch shall control the fluoroscopic device.

S.D. Admin. R. 44:03:01:08.03

26 SDR 96, effective 1/23/2000; 50 SDR 041, effective 10/11/2023

General Authority: SDCL 34-21-4.1, 34-21-15.

Law Implemented: SDCL 34-21-4.1, 34-21-15.