Section XV-762 - Full Inspection Servicing for Teletherapy and Gamma Stereotactic Radiosurgery Units
A. A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during each source replacement and at intervals not to exceed five years to ensure proper functioning of the source exposure mechanism and other safety components.
B. This inspection and servicing shall be performed only by persons specifically licensed to do so by the department, an agreement state, or the U.S. Nuclear Regulatory Commission.
C. A licensee shall maintain a record of the inspection and servicing for the duration of the license. The record shall contain the inspector's name, the inspector's license number, the date of inspection, the manufacturer's name and model number and serial number for both the teletherapy unit and source, a list of components inspected, a list of components serviced and the type of service, a list of components replaced, and the signature of the inspector.

La. Admin. Code tit. 33, § XV-762

Promulgated by the Department of Environmental Quality, Office of Air Quality and Radiation Protection, Radiation Protection Division, LR 18:34 (January 1992), amended by the Office of Environmental Assessment, Environmental Planning Division, LR 26:2590 (November 2000), LR 30:1186 (June 2004), Amended by the Office of the Secretary, Legal Affairs and Criminal Investigations Division, LR 471860 (12/1/2021), Amended by the Office of the Secretary, Legal Affairs Division, LR 4962 (1/1/2023).
AUTHORITY NOTE: Promulgated in accordance with R.S. 30:2001 et seq. and 2104.B.