Performance test | Specifications | Acceptance criteria |
§ 53.62 Full Wind Tunnel Evaluation | Solid VOAG produced aerosol at 2 km/hr and 24 km/hr | Dp50 = 2.5 [MICRO]m ±0.2 [MICRO]m Numerical Analysis Results: 95% [LESS THAN EQUAL TO]Rc [LESS THAN EQUAL TO]105%. |
§ 53.63 Wind Tunnel Inlet Aspiration Test | Liquid VOAG produced aerosol at 2 km/hr and 24 km/hr | Relative Aspiration: 95% [LESS THAN EQUAL TO]A [LESS THAN EQUAL TO]105%. |
§ 53.64 Static Fractionator Test | Evaluation of the fractionator under static conditions | Dp50 = 2.5 [MICRO]m ±0.2 [MICRO]m Numerical Analysis Results: 95% [LESS THAN EQUAL TO]Rc [LESS THAN EQUAL TO]105%. |
§ 53.65 Loading Test | Loading of the clean candidate under laboratory conditions | Acceptance criteria as specified in the post-loading evaluation test (§ 53.62 , § 53.63 , or § 53.64 ). |
§ 53.66 Volatility Test | Polydisperse liquid aerosol produced by air nebulization of A.C.S. reagent grade glycerol, 99.5% minimum purity | Regression Parameters Slope = 1 ±0.1, Intercept = 0 ±0.15 mg, r [GREATER THAN EQUAL TO]0.97. |
40 C.F.R. 53 app Table F-1 to Subpart F of Part 53