Ariz. Admin. Code § tit. 13, ch. 10, art. 1, exh. G-3

Current through Register Vol. 30, No. 50, December 13, 2024
Exhibit G-3 - Standard Calibration Check Procedure Intoxilyzer, Model 8000 (Option P)

THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD QUALITY ASSURANCE PROCEDURES

INTOXILYZER MODEL 8000

STANDARD CALIBRATION CHECK PROCEDURE

(OPTION P)

1.
a. Ensure dry gas tank is attached to instrument and contains a standard alcohol concentration solution alcohol standard.

OR

b. Pour a standard alcohol concentration solution into a clean dry simulator and assemble the simulator.

Ensure that a tight seal has been made. Turn on the simulator and allow temperature to reach 34° C ± 0.2° C

2. Intoxilyzer 8000 display reads "PUSH BUTTON TO START"
3. Go to the "Control Testing Menu". Select "D" for dry control test or "W" for wet control test. After selection is made press ENTER.
4. Air blank completed.
5. Standard Calibration Check completed.
6. Air blank completed.

DPS Form Exh G-3 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. G-3

New Exhibit G-3 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).