Ariz. Admin. Code § tit. 13, ch. 10, art. 1, exh. G-2

Current through Register Vol. 30, No. 45, November 8, 2024
Exhibit G-2 - Standard Calibration Check Procedure, Intoxilyzer Model 8000

THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD QUALITY ASSURANCE PROCEDURES

INTOXILYZER MODEL 8000

STANDARD CALIBRATION CHECK PROCEDURE

QA SPECIALIST ________________________________________ AGENCY ____________________

DATE _______________________________________TIME _________________________________

INTOXILYZER SERIAL # ______________________________LOCATION _______________________

() 1. Ensure that gas tank is attached to instrument and contains a standard alcohol concentration solution _________AC.

OR

Pour a standard alcohol concentration solution _________AC, into a clean dry simulator and assemble the simulator. Ensure that a tight seal has been made. Turn on the simulator and allow temperature to reach 34° C ± 0.2° C

() 2. Intoxilyzer 8000 display reads "PUSH BUTTON TO START"

() 3. Go to the "Control Testing Menu". Select "D" for dry control test or "W" for wet control test. After selection is made press ENTER.

() 4. Air blank completed.

() 5. Calibration check completed. Test results 0._____________AC.

() 6. Air blank completed.

() 7. Remove printed record. Attach the record to the completed checklist.

SIGNATURE _________________________________________________________________________

DPS Form Exh G-2 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. G-2

New Exhibit G-2 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).