034-5 Wyo. Code R. § 5-12

Current through April 27, 2019
Section 5-12 - Office Faculties and Equipment Requirements for Minimal Sedation, Moderate Sedation, Deep Sedation and/or General Anesthesia

(a) Minimal Sedation. Any dentist that administers minimal sedation shall provide the following equipment, which shall be functional and available at all times:

  • (i) A continuous pulse oximeter;
  • (ii) A blood pressure cuff of appropriate size;
  • (iii) Stethoscope or equivalent blood pressure monitoring devices;
  • (iv) An appropriate size bag-valve-mask apparatus or equivalent with an oxygen hook-up; and
  • (v) A manual or automatic external defibrillator.

(b) Moderate Sedation, Deep Sedation and/or General Anesthesia. Any permit holder who administers moderate sedation, deep sedation, and/or general anesthesia shall provide the required equipment listed in subsection (a) and the following additional equipment and faculties, which shall to be functional and available at all times:

  • (i) Suitable operating suite;
  • (ii) Recovery area;
  • (iii) Gas storage facilities with back up tanks and mobile back-up oxygen, which generally meet accepted safety standards;
  • (iv) Suction system;
  • (v) Back-up suction equipment;
  • (vi) Back-up lighting equipment;
  • (vii) Parenteral access or the ability to gain parenteral access, if clinically indicated;
  • (viii) Capnograph (end tidal carbon dioxide monitor);
  • (ix) EKG;
  • (x) Appropriate emergency medications;
  • (xi) Endotracheal tubes suitable for patients being treated;
  • (xii) Endotracheal tube forceps (i.e., magill);
  • (xiii) A laryngoscope with reserve batteries and bulbs;
  • (xiv) Oropharyngeal airways;
  • (xv) Nasopharyngeal airways; and
  • (xvi) At least one (1) additional airway device.

(c) Volatile Anesthesia Delivery Systems. Any permit holder that administers volatile anesthesia shall provide the required equipment listed in subsections (a) and (b) and the following additional equipment and facilities, which shall to be functional and available at all times:

  • (i) Capability to deliver oxygen to a patient under positive pressure, including a back-up oxygen system;
  • (ii) Gas outlets that meet generally accepted safety standards preventing accidental administration of inappropriate gases or gas mixture;
  • (iii) Fail-safe mechanisms for inhalation of nitrous oxide anxiolysis;
  • (iv) The inhalation equipment must have an appropriate scavenging system if volatile anesthetics are used; and
  • (v) Gas storage facilities, which meet generally accepted safety standards.

034-5 Wyo. Code R. § 5-12

Adopted, Eff. 1/8/2015.

Amended, Eff. 9/26/2017.