Registrants shall use only dental X-ray systems and medical X-ray systems for dental imaging that meet the following requirements:
Table 1
Design operating range (kVp) | Measured potential ( kVp) | HVL (millimeters of aluminum equivalent) |
70 and below | 70 and below | 1.5 |
Above 70 | 71 | 2.1 |
80 | 2.3 | |
90 | 2.5 | |
100 | 2.7 |
X1 - X2 <= 0.10 (X1 + X2) |
Where X1 and X2 are the ratios (mR/mAs) for each mA or mAs setting.
(Emax - Emin) <= 0.1E |
Wash. Admin. Code § 246-225A-060
Statutory Authority: RCW 70.98.050 and 70.98.080. 11-19-013, § 246-225A-060, filed 9/7/11, effective 10/8/11; 08-14-074, § 246-225A-060, filed 6/26/08, effective 7/27/08.