Current through Bulletin 2024-20, October 15, 2024
Section R313-35-70 - Radiation Survey InstrumentsSurvey instruments used in determining compliance with R313-15 and R313-35 shall meet the following requirements:
(1) Instrumentation shall be capable of measuring a range from 0.02 millisieverts (2 millirem) per hour through 0.01 sievert (1 rem) per hour.(2) Instrumentation shall be calibrated at intervals not to exceed 12 months and after instrument servicing, except for battery changes.(3) For linear scale instruments, calibration shall be shown at two points located approximately one- third and two-thirds of full-scale on each scale. For logarithmic scale instruments, calibration shall be shown at mid- range of each decade, and at two points of at least one decade. For digital instruments, calibration shall be shown at three points between 0.02 and 10 millisieverts (2 and 1000 millirems) per hour.(4) An accuracy of plus or minus 20 percent of the calibration source shall be demonstrated for each point checked pursuant to R313-35-70(3).(5) The registrant shall perform visual and operability checks of survey instruments before use on each day the survey instrument is to be used to ensure that the equipment is in good working condition. If survey instrument problems are found, the equipment shall be removed from service until repaired.(6) Results of the instrument calibrations showing compliance with R313-35-70(3) and R313-35-70(4) shall be recorded and maintained for a period of three years from the date the record is made.(7) Records demonstrating compliance with R313-35-70(5) shall be made when a problem is found. The records shall be maintained for a period of three years from the date the record is made.Utah Admin. Code R313-35-70