Section 44:67:05:03 - Recovery rooms and services

Each abortion facility shall be designed and equipped for the types of abortion procedures to be performed.

Recovery rooms shall meet the following requirements:

(1) A minimum area of 45 square feet in each recovery station;
(2) Privacy curtains surrounding each recovery station;
(3) A work station and handwashing sink. Handwashing facilities shall be equipped with hand-free controls and have a water supply spout mounted so that the discharge is a minimum of five inches above the rim of the fixture;
(4) Adequate lighting with protective covers to prevent lamps from shattering;
(5) An emergency call system to summon staff; and
(6) Battery operated emergency lighting.

Recovery services shall include:

(1) Provisions for immediate post-procedure care; and
(2) Emergency drugs and supplies specified by the medical staff.