S.C. Code Regs. § § 61-91.2300.2304

Current through Register Vol. 48, No. 11, November 22, 2024
Section 61-91.2300.2304 - Receptacles (II)
A. Recovery Area. Each bed in the recovery area shall have duplex grounding-type receptacles located per NFPA 70, to include one at the head of each bed.
B. Corridors. Duplex receptacles for general use shall be installed approximately 50 feet apart in all corridors and within 25 feet of the ends of corridors.

S.C. Code Regs. § 61-91.2300.2304