25 Pa. Code § 228.74

Current through Register Vol. 54, No.43, October 26, 2024
Section 228.74 - Absorbed dose rate

New equipment shall have a system that provides information from which the absorbed dose rate at a reference point in the treatment volume can be calculated. The radiation detectors specified in § 228.66 (relating to beam monitors) may form part of this system. The dose monitor unit rate shall be displayed at the control panel.

25 Pa. Code § 228.74

The provisions of this §228.74 adopted October 2, 1998, effective 10/3/1998, 28 Pa.B. 4894.