Section 221.33a - Radiation from capacitor energy storage equipment in standby status

Radiation emitted from an X-ray tube when the exposure switch or timer in not activated may not exceed a rate of 2 milliroentgens (0.516 µC/kg) per hour at 5 centimeters from an accessible surface of a fully charged diagnostic source assembly, with the beam-limiting device fully open.