Current through Register Vol. 64, No. 1, January 1, 2025
Section 333-123-0070 - Control Panel FunctionsA control panel must be designed to provide:
(1) An indication of whether electrical power is available at the control panel and if activation of the electronic brachytherapy source is possible;(2) An indication of whether X-rays are being produced;(3) A means for displaying electronic brachytherapy source accelerating voltage(kV), beam current(µA), pre-set value of radiation dose(cGy) and real time display of dose delivered;(4) The means for terminating an exposure at any time; and(5) An access locking control device that can prevent unauthorized use of the electronic brachytherapy device.Or. Admin. Code § 333-123-0070
PH 4-2013, f. & cert. ef. 1-29-13Stat. Auth.: ORS 453.605 - 453.807
Stats. Implemented: ORS 453.605 - 453.807