Current through all regulations passed and filed through October 28, 2024
Section 3701:1-58-96 - Records of periodic spot-checks for remote afterloader units(A) A licensee shall retain a record of each spot-check for remote afterloader units required by rule 3701:1-58-65 of the Administrative Code for three years. (B) The record must include, as applicable: (1) The date of the spot-check; (2) The manufacturer's name, model number, and serial number for the remote afterloader unit and source; (3) An assessment of timer accuracy; (4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and (5) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check. (C) A licensee shall retain a copy of the procedures required by paragraph (B) of rule 3701:1-58-65 of the Administrative Code until the licensee no longer possesses the remote afterloader unit. Ohio Admin. Code 3701:1-58-96
Five Year Review (FYR) Dates: 4/7/2022 and 04/07/2027
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 08/15/2005Five Year Review (FYR) Dates: 06/12/2015 and 06/01/2020
Promulgated Under: 119.03
Statutory Authority: 3743.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 8/15/2005