Section 3701:1-58-23 - Possession, use, and calibration of instruments used to measure the activity of unsealed radioactive material(A) For direct measurements performed in accordance with rule 3701:1-58-25 of the Administrative Code, a licensee shall possess and use instrumentation to measure the activity of unsealed radioactive material before it is administered to each patient or human research subject. (B) A licensee shall calibrate the instrumentation required in paragraph (A) of this rule in accordance with nationally recognized standards or the manufacturer's instructions. (C) A licensee shall retain a record of each instrument calibration required by this rule in accordance with rule 3701:1-58-77 of the Administrative Code. Ohio Admin. Code 3701:1-58-23