In addition to other requirements of this chapter, all diagnostic x-ray systems shall meet the following requirements:
TABLE I | ||||
Half-Value Layer in Millimeters Aluminum | ||||
Design Operating Range (Kilovolts Peak) | Measured Potential (Kilovolts Peak) | Dental Intraoral Manufactured Before Aug. 1, 1974, and on or After Dec. 1, 1980 | Diagnostic X-Ray Systems Manufactured Prior to June 10, 2006 | Diagnostic X-Ray Systems Manufactured on or After June 10, 2006 |
Below 51 | 30 | N/A | 0.3 | 0.3 |
40 | N/A | 0.4 | 0.4 | |
50 | 1.5 | 0.5 | 0.5 | |
51 to 70 | 51 | 1.5 | 1.2 | 1.3 |
60 | 1.5 | 1.3 | 1.5 | |
70 | 1.5 | 1.5 | 1.8 | |
Above 70 | 71 | 2.1 | 2.1 | 2.5 |
80 | 2.3 | 2.3 | 2.9 | |
90 | 2.5 | 2.5 | 3.2 | |
100 | 2.7 | 2.7 | 3.6 | |
110 | 3.0 | 3.0 | 3.9 | |
120 | 3.2 | 3.2 | 4.3 | |
130 | 3.5 | 3.5 | 4.7 | |
140 | 3.8 | 3.8 | 5.0 | |
150 | 4.1 | 4.1 | 5. 4 |
N.D. Admin Code 33.1-10-06-04
General Authority: NDCC 23.1-03-04; S.L. 2017, ch. 199, § 1
Law Implemented: NDCC 23.1-03-03, 23.1-03-04; S.L. 2017, ch. 199, § 18