Operating kVp | Minimum total filtration (Inherent plus added) |
Below 50 kVp | 0.5 mm aluminum |
50-70 kVp | 1.5 mm aluminum |
Above 70 kVp | 2.5 mm aluminum |
Designed operating range (kVp) | Measured operating potential (kVp) | Minimum HVL mm of Al |
Below 51 | 30 | 0.3 |
40 | 0.4 | |
50 | 0.5 | |
51 to 70 | 51 | 1.2 |
60 | 1.3 | |
70 | 1.5 | |
Above 70 | 71 | 2.1 |
80 | 2.3 | |
90 | 2.5 | |
100 | 2.7 | |
110 | 3.0 | |
120 | 3.2 | |
130 | 3.5 | |
140 | 3.8 | |
150 | 4.1 |
0.75 inch (19 mm) of aluminum (pediatric patient - 25 kg.); .......
1.50 inches (38 mm) of aluminum (small adult patient - 50 kg.; .......
1.50 inches (38 mm) of aluminum and 0.02 inch (0.5 mm) of copper (average adult patient -75 kg.); .......
1.50 inches (38 mm) of aluminum and 0.08 inch (2.0 mm) of copper (large adult patient - 100 kg); .......
1.50 inches (38 mm) of aluminum and 0.08 inch (2.0 mm) of copper and 0.12 inch (3.0 mm) of lead (for maximum fluoroscopic exposure rate only). .......
The fluoroscopic exposure rates for the most frequently performed procedure shall be posted so that they are conspicuous to the operator.
N.Y. Comp. Codes R. & Regs. Tit. 10 § 16.58