TABLE 3 | ||||
Computed Tomography Quality Control Requirements | ||||
(To be performed by a licensed radiologic technologist, a qualified medical | ||||
physicist, or a trained service technician) | ||||
Item | Required Test or | Frequency | Standard | |
Procedure | ||||
1. | Equipment Function: | Daily, each | Must work properly | |
Indicators, | day x-rays | |||
Mechanical, and | are taken | |||
other Safety | ||||
Checks. Warm-up | ||||
2. | For film processing, | As specified | As specified in 7:28-22.5 | |
items 2, 5, 7, | in Table 1, | Table 1, Radiographic Quality Control | ||
and 11 QC tests as | Radiographic | Requirements | ||
specified in Table | Quality | |||
1, Radiographic | Control | |||
Quality Control | Requirements | |||
Requirements | ||||
3. | CT Number for Water | Daily | CT equipment or phantom manufacturers' | |
specifications | ||||
4. | Field Uniformity | Daily | CT equipment or phantom manufacturers' | |
specifications | ||||
5. | Laser Film Printer | Weekly | Recommended control | *OD=Optical |
Quality Control | limits | Density | ||
SMPTE Test Pattern | Inverted gray | |||
scale | ||||
0% patch 2.45 +/- | 0% patch 2.50 + | |||
0.15 OD* | 0.15 OD | |||
10% patch 2.10 +/- | 10% patch 2.25 +/- | |||
0.15 OD | 0.15 OD | |||
40% patch 1.15 +/- | 40% patch 1.35 +/- | |||
0.15 OD | 0.15 OD | |||
90% patch 0.30 +/- | 90% patch 0.30 +/- | |||
0.08 OD | 0.08 OD | |||
*The 5% patch should just be visible | ||||
inside of the 0% patch. | ||||
The 95% patch should be visible inside | ||||
the 100% patch. | ||||
6. | Low Contrast | Initially and | CT equipment or phantom manufacturers' | |
Resolution | Monthly | specifications | ||
7. | High Contrast | Initially and | CT equipment or phantom manufacturers' | |
Spatial Resolution | Monthly | specifications | ||
8. | Noise | Initially and | CT equipment or phantom manufacturers' | |
Monthly | specifications | |||
9. | Table Position | Initially and | +/- 2 mm | |
Indicator Accuracy | Monthly | |||
10. | Scan Increment | Initially and | +/- 1 mm | |
Accuracy | Monthly | |||
11. | Scan Localization | Initially and | +/- 5 mm | |
Light Accuracy | Monthly | |||
12. | Medical Physicist's | Initially and | As required in 7:28-22.10 | |
QC Survey | annually | |||
13. | Quality Assurance | Initially and | As required in 7:28-22.4(a)7 | |
Program Review | annually |
N.J. Admin. Code § 7:28-22.7