Section 7:28-16.6 - Panoramic radiographic installations
(a) No person shall use any panoramic radiographic unit or cause it to be used unless the following requirements are met:
1. The x-ray field in the plane of the image receptor shall not exceed each dimension of the image receptor by more than two percent of the SID, when the axis of the x-ray beam is perpendicular to the plane of the image receptor. In addition, the center of the x-ray field shall be aligned with the center of the image receptor to within two percent of the SID or there shall be a device provided to both size and align the x-ray field such that the x-ray field at the plane of the image receptor does not extend beyond any edge of the image receptor.
2. These units shall meet all the requirements of this subchapter with the exception of 7:28-16.3(a)3 and 7:28-16.5.