Current through Register Vol. 49, No. 23, December 2, 2024
Section 19 CSR 20-10.140 - Special Requirements for Medical Radiographic InstallationsPURPOSE: This rule establishes special requirements for medical radiographic X-ray installations.
(1) A device shall be provided to terminate the exposure after a preset time or exposure.(2) The exposure switch shall be so arranged that it cannot be conveniently operated outside a shielded area. Exposure switches for spot film devices used in conjunction with fluoroscopic tables are excepted from this requirement. AUTHORITY: section 192.420, RSMo 1986.* This rule was previously filed as 13 CSR 50-90.140. Original rule filed Nov. 9, 1964, effective Dec. 9, 1964. *Original authority: 192.420, RSMo 1963.