7. Linearity, Uncertified X-Ray Systems Only. The following requirements apply when the equipment is operated on a power supply as specified by the manufacturer for any fixed x-ray tube potential within the range of 40 percent to 100 percent of the maximum rated. aEquipment having independent selection of x-ray tube current (mA). The average ratios of exposure to the indicated milliampere-seconds product mR/mAs (C/kg/mAs) obtained at any two tube current settings shall not differ by more than 0.10 times their sum. This is:Click Here To View Image
where
X1 and X2are the average mR/mAs (C/kg/mAs) values obtained at any two tube current settings.
bEquipment having a combined x-ray tube current-exposure time product (mAs) selector, but not a separate tube current (mA) selector. The average ratios of exposure to the indicated milliampere-seconds product mR/mAs (C/kg/mAs) obtained at any two mAs selector settings shall not differ by more than 0.10 times their sum. This is:Click Here To View Image
where X1 and X2 are the average mR/mAs (C/kg/mAs) values obtained at any two mAs selector settings.
c. Measuring compliance. Determination of compliance shall be based on 4 exposures, of no less than 0.05 seconds each, taken within a time period of one hour, at each of the two settings. These two settings may include any two focal spot sizes provided that neither focal spot size is equal to or less than 0.45 millimeter, in which case the two settings shall be restricted to the same focal spot size. For purposes of this requirement, focal spot size is the nominal focal spot size specified by the x-ray tube manufacturer.