In addition to other requirements of this section, all diagnostic x-ray systems shall meet the following requirements:
3 Leakage Radiation from the Diagnostic Source Assembly. The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source shall not exceed 100 milliroentgens (25.8 µC/kg) in 1 hour when the x-ray tube is operated at its leakage technique factors. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters. If leakage technique factors cannot be set on the control panel, then compliance shall be determined by measuring leakage at maximum kVp and an appropriate mAs.
4 Radiation from Components Other Than the Diagnostic Source Assembly The radiation emitted by a component other than the diagnostic source assembly shall not exceed 2 milliroentgens (0.516 µC/kg) in 1 hour at 5 centimeters from any accessible surface of the component when it is operated in an assembled x-ray system under any conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
5 Beam Quality.
TABLE I
X-ray tube voltage | (kilovolt peak) | Minimum HVL (mm of Al) | ||
Designed Operating Range | Measured Operating Potential | Dental Systems | Medical X-ray Systems Manufactured before 06/10/2006 | Medical X-ray Systems Manufactured after 06/10/2006 |
Below 50 | 30 | 1.5 | 0.3 | 0.3 |
40 | 1.5 | 0.4 | 0.4 | |
50 | 1.5 | 0.5 | 0.5 | |
50 to 70 | 51 | 1.5 | 1.3 | 1.3 |
60 | 1.5 | 1.5 | 1.5 | |
70 | 1.5 | 2.1 | 1.8 | |
Above 70 | 71 | 2.1 | 2.1 | 2.5 |
80 | 2.3 | 2.3 | 2.9 | |
90 | 2.5 | 2.5 | 3.2 | |
100 | 2.7 | 2.7 | 3.6 | |
110 | 3.0 | 3.0 | 3.9 | |
120 | 3.2 | 3.2 | 4.3 | |
130 | 3.5 | 3.5 | 4.7 | |
140 | 3.8 | 3.8 | 5.0 | |
150 | 4.1 | 4.1 | 5.4 |
15 Miss. Code. R. 21-78-1.6.4