Open-beam configurations must have a readily discernible indication of:
An easily visible warning light labeled with the words "X-RAY ON" or other visible warning indicator that clearly shows the equipment is producing ionizing radiation, must be located near a switch that energizes an x-ray tube and illuminated only when the tube is energized.
Warning devices must be labeled so that their purpose is easily identified.
Minn. R. agency 144, ch. 4732, RADIATION THERAPEUTIC REQUIREMENTS, pt. 4732.1063
Statutory Authority: MS s 144.12