16-163-17 Me. Code R. § 2

Current through 2024-51, December 18, 2024
Section 163-17-2 - Airway management supplies

Click to view image

Click to view image

Click to view image

1 Automatic, pressure cycled resuscitators are not acceptable.

2 CPAP Device Requirements: Full face mask, continuous flow device, capable of adjusting FiO2, capable of regulating Positive End Expiratory Pressure (PEEP), latex-free, and the ability to attach a nebulizer.

3 Hyperangulated blades are not accepted

4 Suitable for use with adult and pediatric patients

5 Portable unit to provide pharyngeal suction of at least 11,8 inches mercury (300mm Hg) within 4 seconds after the suction tube is clamped closed. Unit must have trap bottle, and be equipped with rigid pharyngeal suction tip and appropriate catheters

6 Capable to provide pharyngeal suction of at least 11,8 inches mercury (300mm Hg) within 4 seconds after the suction tube is clamped closed. Unit must have trap bottle and be equipped with rigid pharyngeal suction tip and appropriate catheters. The unit must be electrically powered - capable of operating from its own (internal) battery.

7 May be commercially prepared, but must contain items listed

8 MDPB approved percutaneous cricothyroidotomy kits only if they follow the method of piercing the cricothyroid membrane

9 recommended 6.0 internal diameter for adults

10 Or other FDA Approved Patient Antiseptic Skin Preparations as a means to prep and cleanse the skin

11 May be commercially prepared, but must contain items listed

12 It is recommended to have NO MORE THAN one device per class (periglottic and transglottic).

13 It is recommended to have NO MORE THAN one device per class (periglottic and transglottic).

16-163 C.M.R. ch. 17, § 2