10-144-220 Me. Code R. § G-L-2432

Current through 2024-51, December 18, 2024
Section 144-220-G-L-2432 - Records of calibration measurements of brachytherapy sources
A. A licensee shall maintain a record of the calibrations of brachytherapy sources required by G.432 for three years after the last use of the source.
B. The record must include:
(1) The date of the calibration;
(2) The manufacturer's name, model number, and serial number for the source and the instruments used to calibrate the source;
(3) The source output or activity;
(4) The source positioning accuracy within the applicators; and
(5) The signature of the authorized medical physicist.

10-144 C.M.R. ch. 220, § G-L-2432