Section 144-220-G-F-457 - Therapy-related computer systems

The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

A. The source-specific input parameters required by the dose calculation algorithm;
B. The accuracy of dose, dwell time, and treatment time calculations at representative points;
C. The accuracy of isodose plots and graphic displays; and
D. The accuracy of the software used to determine sealed source positions from radiographic images.