10-144-220 Me. Code R. § G-C-61

Current through 2024-51, December 18, 2024
Section 144-220-G-C-61 - Calibration of survey instruments
A. A licensee shall calibrate the survey instruments used to show compliance with this Part and Part D before first use, annually, and following a repair that affects the calibration. A licensee shall:
(1) Calibrate all required scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;
(2) Have each radiation survey instrument calibrated:
(a) At energies appropriate for use and intervals not to exceed 12 months or after instrument servicing, except battery changes;
(b) For linear scale instruments, at two points located approximately one- third and two-thirds of full-scale on each scale; for logarithmic scale instruments, at mid-range of each decade, and at two points of at least one decade; and for digital instruments, at three points between 0.02 and 10 millisieverts (2 and 1000 mrem) per hour; and
(c) For dose rate instruments, so that an accuracy within plus or minus 20 percent of the true radiation dose rate can be demonstrated at each point checked.; and
(3) Conspicuously note on the instrument the apparent dose rate from a dedicated check source as determined at the time of calibration, and the date of calibration.
B. A licensee may not use survey instruments if the difference between the indicated exposure rate and the calibrated exposure rate is more than 20 percent.
C. A licensee shall check each survey instrument for consistent response with a dedicated check source before each use.
D. A licensee shall retain a record of each survey instrument calibration in accordance with G.2061

10-144 C.M.R. ch. 220, § G-C-61