La. Admin. Code tit. 51 § VII-313

Current through Register Vol. 50, No. 11, November 20, 2024
Section VII-313 - Pasteurization Equipment Tests, Examinations and Sealing
A. The state health officer shall perform the tests using the methodology prescribed in the PMO on the instruments and devices of each pasteurizer in each dairy plant and frozen dessert manufacturing plant indicated in the table below initially upon installation; and at least once each three months, including the remaining days of the month in which the equipment tests are due and whenever any alteration or replacement is made which may affect the proper operation of the instrument or device. Provided, that the holding time test shall be conducted at least once each six months, including the remaining days of the month in which the equipment test is due. A copy of the test report shall be retained by the plant.

Test No.

Equipment/Device/ Instrument

Test Type

1

Vat, HTST, HHST, aseptic indicating and airspace thermometers

Temperature accuracy

2

Vat, HTST, HHST, aseptic recording thermometer

Temperature accuracy

3

Vat, HTST, HHST, aseptic recording thermometer

Time accuracy

4

Vat, HTST, HHST, aseptic indicating and recording thermometer

Recording vs Indicating thermometer

5.1

HTST, HHST FDD

Leakage pass FDD

5.2

HTST, HHST FDD

FDD freedom of movement

5.3

HTST, HHST FDD

Device assembly (single stem)

5.4

HTST, HHST FDD

Device assembly (dual stem)

5.5

HTST FDD

Manual diversion

5.6

HTST, HHST FDD

Response time

5.7

HTST, HHST FDD

Time delay (inspect)

5.8

HTST, HHST FDD

Time delay (CIP)

5.9

HTST FDD

Time delay (leak detect flush)

6

Vat leak protector valve(s)

Leakage

7

HTST indicating thermometers

Response time

8

HTST recording thermometers

Response time

9.1

HTST pressure switches

Regenerator pressures

9.2.1

HTST, HHST, aseptic differential pressure controllers

Calibration

9.2.2

HTST differential pressure controllers

Regenerator pressure

9.2.3

HHST and aseptic differential pressure controllers

Regenerator pressure

9.3.1

HTST booster pump/FDD

Inter-wiring check

9.3.2

HTST booster pump/metering pump

Inter-wiring check

10.1

HTST FDD

Temperature cut-in/cut-out

10.2

HHST FDD, aseptic divert system (indirect heat)

Temperature cut-in/cut-out

10.3

HHST FDD, aseptic divert system (direct heat)

Temperature cut-in/cut-out

11.1

HTST holding tubes/timing pumps (except meter based)

Holding time

11.2.a

HTST holding tubes/magnetic flow meter based timing systems

Holding time

11.2.b

HTST, HHST, aseptic magnetic flow meter based timing systems

Flow alarm

11.2.c

HTST, HHST, aseptic magnetic flow meter based timing systems

Loss of signal/low flow

11.2.d

HTST magnetic flow meter based timing systems

Flow rate cut-in/cut-out

11.2.e

HTST magnetic flow meter based timing systems

Time delay

11.3

HHST holding tubes indirect heat

Holding time

11.4

HHST holding tubes direct injection heat

Holding time

11.5

HHST holding tubes direct infusion heat

Holding time

12.1

HHST, aseptic systems indirect heating

Sequence logic

12.2

HHST, aseptic systems direct heating

Sequence logic

13

HHST, aseptic systems

Pressure in the holding tubes

14

HHST, aseptic systems using direct injection heating

Pressure differential across injector

15

Vat, HTST, HHST, Aseptic (all electronic controls)

Electro-Magnetic Interference

B. Plants being regulated under the provisions of Chapter 11 (Dairy Plant HACCP System) shall be responsible for the performance for all above required tests should the state health officer fail to perform them at the required frequency.
C. The state health officer shall affix regulatory seals to all pasteurization equipment, as prescribed by the PMO, after testing the equipment.
D. The state health officer shall provide the plant a copy of the pasteurization equipment test report.
E. The plant shall notify the state health officer immediately if the regulatory seals are broken.
F. The pasteurization equipment shall not be operated without authorization from the state health officer when any of the required regulatory seals are broken.

La. Admin. Code tit. 51, § VII-313

Promulgated by the Department of Health and Hospitals, Office of Public Health, LR 37:2653 (September 2011).
AUTHORITY NOTE: Promulgated in accordance with the provisions of R.S. 40:4(A)(1)(a). Also see R.S. 40:5(2)(3)(5)(7) (15)(17) and R.S. 40:922.