La. Admin. Code tit. 33 § XV-908

Current through Register Vol. 50, No. 11, November 20, 2024
Section XV-908 - Particle Accelerator Controls and Interlock Systems
A. Instrumentation, readouts, and controls on the particle accelerator control console shall be clearly identified, legible, and easily discernible.
B. All entrances into a target room or other high radiation area shall be provided with interlocks that shut down the machine under conditions of barrier penetration or intrusion.
C. When an interlock system has been tripped, it shall be possible to resume operation of the accelerator only by manually resetting controls at the position where the interlock has been tripped and, lastly, at the main control console.
D. Each radiation safety interlock shall be on a circuit that shall allow its operation independently of all other safety interlocks.
E. All radiation safety interlocks shall be fail-safe, i.e., designed so that any defect or component failure in the interlock system prevents operation of the accelerator.
F. A "scram button" or other emergency power cutoff switch shall be located and easily identifiable in all high radiation areas. Such a cutoff switch shall include a manual reset so that the accelerator cannot be restarted from the accelerator control console without resetting the emergency power cutoff switch.

La. Admin. Code tit. 33, § XV-908

Promulgated by the Department of Environmental Quality, Nuclear Energy Division, LR 13:569 (October 1987), amended by the Office of Air Quality and Radiation Protection, Radiation Protection Division, LR 19:1421 (November 1993).
AUTHORITY NOTE: Promulgated in accordance with R.S. 30:2001 et seq.