Current through Register Vol. 49, No. 2, January 10, 2025
Section 265.2950 - Emergency Electric Servicea) An emergency source of electricity shall be provided.b) Birth centers shall be permitted to use a battery system for emergency power. The following is required: 1) Illumination of means of egress as required in the Life Safety Code (NFPA 101);2) Illumination of birth and recovery rooms;3) Illumination of exit and exit directional signs;4) Fire alarm and alarms required for nonflammable medical gas systems, if nonflammable medical gas systems are installed; and5) Type 3 emergency electrical service that meets all NFPA 99 requirements of this type of system.Ill. Admin. Code tit. 77, § 265.2950