Rule 25-F407 - MAINTENANCE - TANNING EQUIPMENT, TIMER, GOOD REPAIR AND ACCURATE CALIBRATION*407.1Tanning equipment and devices shall be maintained in good repair in order to prevent any mechanical safety hazards.
407.2The licensee shall ensure that tests are performed quarterly on each assembly of tanning equipment or device and documented in writing to ensure that:
(a) The timer is accurately calibrated to within ten percent (10%) of the maximum timer interval for the product as specified in Section 402.3; and(b) The consumer is able to terminate radiation emissions manually as specified in Section 400.4.407.3Maintenance tests shall include the date of the test, and the timer test shall include the indicated time versus the measured time.
407.4The timer shall be tested at the tanning equipment manufacturer's recommended maximum exposure time.
D.C. Mun. Regs. tit. 25, r. 25-F407
Final Rulemaking published at 60 DCR 3582 (March 15, 2013)Authority: Section 4902 of the Department of Health Functions Clarification Act of 2001 (Act), effective October 3, 2001 (D.C. Law 14-28; D.C. Official Code § 7-731(a)(8) (2008 Repl. & 2012 Supp.)) and Mayor's Order 2007-63, dated March 8, 2007.