Rule 22-B10721 - DISTRIBUTION
10721.1

Each manufacturer shall establish and maintain procedures for control and distribution of finished devices to ensure that only those devices approved for release are distributed and that purchase orders are reviewed to ensure that ambiguities and errors are resolved before devices are released for distribution. Where a device's fitness for use or quality deteriorates over time, the procedures shall ensure that expired devices or devices deteriorated beyond acceptable fitness for use are not distributed.

10721.2

Each manufacturer shall maintain distribution records which include or refer to the location of:

(a) The name and address of the initial consignee;
(b) The identification and quantity of devices shipped;
(c) The date shipped; and
(d) Any control number(s) used.

D.C. Mun. Regs. tit. 22, r. 22-B10721

Final Rulemaking published at 60 DCR 10252 (July 12, 2013)
Authority: Section 19(a)(3) of the District of Columbia Pharmacist and Pharmacy Regulation Act of 1980, effective September 16, 1980 (D.C. Law 3-98; D.C. Official Code § 47-2885.18(a)(3) (2005 Repl.)); Mayor's Order 98-48, dated April 15, 1998; Section 4902 of the Fiscal Year 2002 Budget Support Act of 2001, effective October 3, 2001 (D.C. Law 14-28; D.C. Official Code § 7-731 (2008 Repl.)); Section 15 of the District of Columbia Medical Device Manufacture and Distribution Licensure Act of 1990, effective June 13, 1990 (D.C. Law 8-137; D.C. Official Code § 48-714(a) (2005 Repl.)); and Mayor's Order 98-88, dated May 29, 1998.