Del. Admin. Code tit. 16, Department of Health and Social Services, Division of Public Health, Health Systems Protection (HSP), 4465, pt. K, app A

Current through Register Vol. 28, No. 7, January 1, 2025
Appendix A - Violation Classification (Typical/Not all inclusive)

Radiation Source Facility

Severity Level 1Severity Level 2
1. Operating without a permit 1. Registration Form
2. Personnel Overexposure 2. Notice to Employees
3. Radiation Levels Excessive 3. Safety Procedures
4. Personnel Monitoring 4. Personnel Monitoring Records

Medical Uses

Severity Level 1Severity Level 2
1. Patient exposure (ESE) 12.
2. Tube support 13. Maintenance of x-ray system records
3. Technique indicators 14. Speed of film/screen
4. Multiple tube indication 15. Technique chart
5. Gonadal shielding 16. Warning label
6. Operator apron/barrier 17. Patient dose of 1 gray (100 rads) or more not reported to the Agency

Radiographic

Severity Level 1Severity Level 2
1. Location of x-ray controls 1. Visual/audio signal
2. Position indicating device 2. Adjustment of x-ray field
3. Beam Collimation 3. Indication of field size, upon adjustments
4. Filtration deficiency of 0.2+ mm 4. Means to limit source to skin distance
5. Variation in timer linearity of 15% or more 5. Filtration deficiency of 0.2 mm or less
6. Variation in exposure reproducibility of 15% or more 6. Variation in timer reproducibility linearity more that 10% but less than 15%
7. Total misalignment of x-ray/light field edges of 5% or more 7. Variation in exposure reproducibility more than5% but less than 10%
8. Total misalignment of x-ray beam/image receptor centers of 5% or more 8. Total misalignment of light/x-ray field edges more than 2% but less than 5%
9. Discorrespondence of indicated x-ray field with beam limited x-ray field of 5% or more 9. Total misalignment of x-ray beam/image receptor centers more than 2% but less than 5%
10. Discorrespondence of indicated x-ray with beam limited x-ray field more than 2% but less than 5%

Fluoroscopic

Severity Level 1Severity Level 2
1. Activation of x-ray production 1. Posting of exposure rate measurements
2. Annual exposure rate measurement 2. Measurement records and posting of same
3. Exposure rate 3. Adjustment of field
4. Useful beam protective barrier 4. Means of field adjustment
5. Limitation of x-ray field to image receptor 5. Indication of kV and mA
6. Minimum field at maximum SID 6. Means to indicate that beam is perpendicular to image receptor
7. High level control exposure rate limit 7. Audible signal to reset
8. Timing device 8. Minimum source to skin distance
9. Control of scattered radiation 9. Exposure rate due to transmission through primary barrier
10. Failure to report patient dose of 1+ gray to the Agency

Industrial Radiographic

Severity Level 1Severity Level 2
1. > 2 mR/hr. Failure to maintain radiation level from device < 4" in diameter below maximum allowable limit of 50 mR/hr. at 6" < 2 mR/hr.
2. > 2 mR/hr. Failure to maintain radiation level from device > 4" diameter below maximum allowable limit of 200 mR/hr. at surface < 2 mR/hr.
3. > 2 mR/hr. Failure to maintain radiation level from storage container below maximum allowable limit of 200 mR/hr. at the surface < 2 mR/hr.
4. > 2 mR/hr. Failure to maintain radiation level from storage container below maximum allowable limit of 10 mR/hr. at a distance of 1 meter < 2 mR/hr.
Locking of Source
5. Radiation Source Storage Container was found unlocked and not under the direct surveillance of authorized individual
6. Exposure device and/or storage container was not locked during transit or prior to unit being made secure
7. Device was not physically secured to prevent tampering or removal

Survey Instruments

Severity Level 1Severity Level 2
1. Failure to adequately maintain survey instruments 1. Failure to possess calibrated back up instrument
2. Survey meter not capable of detecting 2-1000 mR/hr. 2. Failure to conduct quarterly calibrations
3. Failure to calibrate unit after servicing
4. Failure to achieve accuracy of (+ 20%)
5. Failure to calibrate each scale using 2 points other than zero
6. Failure to maintain calibration records for two years

Leak Testing/Repair

Severity Level 1Severity Level 2
1. Service to sealed source not conducted by authorized individual 1. Failure to conduct leak test within 6 month interval
2. Failure to immediately withdraw contaminated unit from service to effect decontamination/repair 2. Leak test procedure not capable of detecting unit 0.005 uCi of removable contamination
3. Failure to properly label unsecured source 3. Failure to record test results in microcuries
4. Failure to notify agency within 5 days of receiving test results

Inventory/Inspection

Severity Level 1Severity Level 2
1. Failure to perform quarterly operational safety maintenance and inspection 1. Failure to conduct quarterly inventory
2. Failure to withdraw defective unit from service 2. Failure to maintain utilization log
3. Failure to conduct and record alarm tests at the beginning of each period of use 3. Failure to log required information
4. Failure to properly document quarterly inspection and maintenance of radiation devices
5. Failure to record and maintain test results

Limitations/Procedures/Controls

Severity Level 1Severity Level 2
1. Individual is not qualified Industrial Radiographer 1. Failure to maintain training and test records
2. Individual is not a qualified radiographer's assistant trainee 2. Failure to record dosimetry data
3. Qualified operator(s) not wearing personal monitoring devices
4. Operating and Emergency procedures do not meet the requirements

Precautionary Procedures

Severity Level 1Severity Level 2
1. Failure of qualified operators to maintain direct surveillance 1. Failure to post required information
2. Failure to have calibrated and operable survey instrument available 2. Failure to maintain survey records for 2 years
3. Failure to conduct physical radiation survey
4. Failure to maintain records at temporary job site

Special Requirements

Severity Level 1Severity Level 2
1. Failure to meet requirements for enclosed radiography 1. Failure to maintain evaluation records for 2 years
2. Failure to meet exemption requirements and maintain unit

Analytical

Severity Level 1Severity Level 2
1. Failure to provide safety device 1. Failure to label equipment
2. Failure to provide warning devices 2. Failure to have appropriate warning lights
3. Failure to install shutter 3. Failure to provide protective cabinet to prevent leakage
4. Failure to construct tube housing in manner which ensures compliance 4. Failure to conduct Radiation Surveys
5. Failure to obtain Radiation Safety Officer's approval for by-passing safety device 5. Failure to document survey record
6. Failure to provide and/or ensure the use of extremity dosimeters 6. Failure to post area
7. Failure to instruct and/or ensure operator competence 7. Failure to secure unused ports

Particle Accelerators Limitations

Severity Level 1Severity Level 2
1. Operator did not receive radiation safety instruction or could not demonstrate understanding of radiation safety 1. Operator had not received copies
2. Operator unable to demonstrate competency in the use of the accelerator 2. Radiation Safety Committee/Radiation Safety Officer is not authorized to terminate operations

Controls/Interlocks

Severity Level 1Severity Level 2
1. HRA not provided with Interlock 1. Instrumentation and controls are not clearly identified
2. Interlock does not require manual reset 2. Scram buttons do not require manual reset
3. Safety interlocks not independently wired 3. Not all HRA entrances are equipped with warning lights
4. All safety interlocks are not fail safe 4. Not all HRA are equipped with audible warning devices which activate for 15 seconds
5. Scram buttons not located in HRA 5. HRA barriers/pathways are not identified

Operating Procedures

Severity Level 1Severity Level 2
1. Particle accelerator not unsecured from unauthorized use
2. Bypass of safety interlock not authorized by Radiation Safety Officer and/or Radiation Safety Committee or not recorded and posted 2. Safety interlocks are used as routine "off" switch
3. Electrical circuit diagram not maintained and available
4. Current operating and emergency procedures are not kept at control panel
5. Record of quarterly safety device operability check are not available

Radiation Monitoring

Severity Level 1Severity Level 2
1. Radiation protection survey not performed and/or documented by an approved person following operation or facility changes 1. Portable monitoring equipment not available, operable and calibrated)
2. Radiation levels not continuously monitored in all HRAs 2. All area monitors are not calibrated annually
3. Periodic surveys of airborne particulates are not performed
4. Periodic Smear Surveys are not conducted for contamination
5. All area surveys are not performed according to proper written procedures
6. Current records of all surveys and tests were not available at facility

Ventilation Systems

Severity Level 1
1. Means are not provided to ensure compliance with Part D airborne concentrations
2. Airborne concentrations in excess of Part D limits were discharged to an uncontrolled area

Notice and Reports to Workers

Posting of Notices

Severity Level 1Severity Level 2
1.
2. Facility radiation permit/document not posted
3. Radiation operating procedures not posted
4. Notice of violation for radiological working conditions not posted
5. Notice to employees, Agency Form X not posted
6. Violations not posted within five (5) days nor remain posted for at least (5) days
7. Documents do not appear in sufficient number of conspicuous places

Instructions to Workers

Severity Level 1Severity Level 2
1. Workers were not instructed to report any conditions that could cause unnecessary exposure to radiation 1. Workers were not kept informed of radiation sources
2. Workers were not instructed about warnings for unusual occurrence or malfunction that may involve exposure to radiation 2. Workers were not instructed of health protection problems associated with radiation exposure
3. Workers were not advised of radiation exposure reports 3. Workers were not instructed to observe applicable parts of DRCR

Notifications

Severity Level 1Severity Level 2
1. Written report of specified Radiation exposure data are not given to the worker 1. Workers are not advised annually of their radiation exposure
2. Workers are not furnished a radiation exposure report within 30 days after the licensee becomes informed of exposure or termination of employment 2. As required pursuant to Part D, Sections 46.0, 47.0 and 48.0 exposed individuals are not provided a report of their radiation exposure

Representatives of Registrant/Workers

Severity Level 1
1. Agency not afforded opportunity to inspect equipment and activities
2. Agency not permitted to consult with workers privately
3. Worker authorized representative not given opportunity to accompany agency during inspection of physical working conditions
4. Worker representative does not meet qualifications set forth in routine radiation
5. Different facility/worker representatives not permitted to accompany agency on inspection
6. Mutually agreed upon "outside" individual(s) was not permitted to accompany agency inspectors
7. Worker(s) was not allowed to privately consult with the Agency inspector about perceived radiological condition
8. A worker has been discharged or discriminated against for filing a radiological complaint on behalf of himself or others

Therapeutic Radiation Machines

Severity Level 1Severity Level 2
1. All non-compliance with the requirements of Part X are Severity Level 1 violations.

Del. Admin. Code tit. 16, Department of Health and Social Services, Division of Public Health, Health Systems Protection (HSP), 4465, pt. K, app A

16 DE Reg. 1274 (6/1/2013)
22 DE Reg. 401( 11/1/2018) (Final)