42 C.F.R. § 84.90

Current through October 31, 2024
Section 84.90 - Breathing resistance test; inhalation
(a) Resistance to inhalation airflow will be measured in the facepiece or mouthpiece while the apparatus is operated by a breathing machine as described in § 84.88 .
(b) The inhalation resistance of open-circuit apparatus shall not exceed 32 mm. (1.25 inch) water-column height (at a flow rate of 120 liters per minute).
(c) The inhalation resistance of closed-circuit apparatus shall not exceed the difference between exhalation resistance (§ 84.91(e) ) and 10 cm. (4 inches) water-column height.

42 C.F.R. §84.90