This section describes two optional methods, using propane as a tracer gas, to verify CVS and PFD flow streams. You may use good engineering judgment and safe practices to use other tracer gases, such as CO2 or CO. The first method, described in paragraphs (a) through (e) of this section, applies for the CVS diluted exhaust flow measurement system. The first method may also apply for other single-flow measurement systems as described in Table 2 of § 1065.307 . Paragraph (g) of this section describes a second method you may use to verify flow measurements in a PFD for determining the PFD dilution ratio.
TABLE 1 OF § 1065.341 -TROUBLESHOOTING GUIDE FOR PROPANE CHECKS
Problem | Recommended corrective action |
Incorrect analyzer calibration | Recalibrate, repair, or replace the FID analyzer. |
Leaks | Inspect CVS tunnel, connections, fasteners, and HC sampling system. Repair or replace components. |
Poor mixing | Perform the verification as described in this section while traversing a sampling probe across the tunnel's diameter, vertically and horizontally. If the analyzer response indicates any deviation exceeding ±2% of the mean measured concentration, consider operating the CVS at a higher flow rate or installing a mixing plate or orifice to improve mixing. |
Hydrocarbon contamination in the sample system | Perform the hydrocarbon-contamination verification as described in § 1065.520 . |
Change in CVS calibration | Perform a calibration of the CVS flow meter as described in § 1065.340 . |
Flow meter entrance effects | Inspect the CVS tunnel to determine whether the entrance effects from the piping configuration upstream of the flow meter adversely affect the flow measurement. |
Other problems with the CVS or sampling verification hardware or software | Inspect the CVS system and related verification hardware, and software for discrepancies. |
40 C.F.R. §1065.341