Current through November 30, 2024
Section 213.94 - [Effective 12/18/2025] Medical gas containers and closures(a) Medical gas containers and closures shall not be reactive, additive, or absorptive so as to alter the safety, identity, strength, quality, or purity of the gas beyond the official or established requirements.(b) Container closure systems shall provide adequate protection against foreseeable external factors in storage and use that can cause deterioration or contamination of the medical gas.(c) Medical gas containers and closures shall be clean to assure that they are suitable for their intended use.(d) Standards or specifications, methods of testing, and, where indicated, methods of cleaning shall be written and followed for medical gas containers and closures.(e) Medical gas containers and closures must meet the following requirements-(1)Gas-specific use outlet connections. Portable cryogenic medical gas containers that are not manufactured with permanent gas use outlet connections (e.g., those that have been silver-brazed) must have gas-specific use outlet connections that are attached to the valve body so that they cannot be readily removed or replaced (without making the valve inoperable and preventing the containers' use) except by the manufacturer. For the purposes of this paragraph (e)(1), the term manufacturer includes any individual or firm that fills high-pressure medical gas cylinders or cryogenic medical gas containers. For the purposes of this section, a portable cryogenic medical gas container is one that is capable of being transported and is intended to be attached to a medical gas supply system within a hospital, healthcare entity, nursing home, other facility, or home healthcare setting, or is used to fill small cryogenic gas containers for use by individual patients. The term excludes cryogenic containers that are not designed to be connected to a medical gas supply system, e.g., tank trucks, trailers, rail cars, or small cryogenic gas containers for use by individual patients (including portable liquid oxygen units as defined at § 868.5655 of this chapter).(2)Gauges for certain medical gas containers. Portable cryogenic medical gas containers as described in paragraph (e)(1) of this section and small cryogenic gas containers for use by individual patients (including portable liquid oxygen units as defined at § 868.5655 of this chapter) must have a working gauge sufficient to assist the user in determining whether the container contains an adequate supply of medical gas for continued use.(3)Label and coloring requirements. The labeling specified at § 201.328(a) of this chapter must be affixed to the container in a manner that does not interfere with other labeling. Each such label as well as materials used for coloring medical gas containers must be reasonably resistant to fading, durable when exposed to atmospheric conditions, and not readily soluble in water.