10 C.F.R. § 35.2432

Current through September 30, 2024
Section 35.2432 - Records of calibration measurements of brachytherapy sources
(a) A licensee shall maintain a record of the calibrations of brachytherapy sources required by § 35.432 for 3 years after the last use of the source.
(b) The record must include-
(1) The date of the calibration;
(2) The manufacturer's name, model number, and serial number for the source and the instruments used to calibrate the source;
(3) The source output or activity;
(4) The source positioning accuracy within the applicators; and
(5) The name of the individual, the source manufacturer, or the calibration laboratory that performed the calibration.

10 C.F.R. §35.2432

67 FR 20370, Apr. 24, 2002, as amended at 68 FR 19326, Apr. 21, 2003