Current through Register Vol. 30, No. 50, December 13, 2024
Section R9-7-608 - Mobile Diagnostic Radiographic and Mobile Fluoroscopic Systems, Except Dental Panoramic, Cephalometric, Dental CT, or Dental Intraoral Radiographic SystemsA. Equipment 1. All requirements of R9-7-607(A) and (B) apply.2. For mobile radiographic units the registrant shall provide a "dead-man" switch, together with an electrical cord of sufficient length so that the operator can stand out of the useful beam and at least 1.82 meters (6 feet) from the patient during all x-ray exposures.3. A registrant shall ensure that a cone, spacer frame, or inherent provision is made so that the equipment is not operated at source-skin distances of less than 20.3 centimeters (8 inches).B. Structural shielding. If a mobile unit is used routinely in one location, it is considered a fixed installation subject to the shielding requirements in R9-7-603(C), and R9-7-607(C).C. Operating procedures 1. All provisions of R9-7-607(D) apply.2. An individual who operates a mobile x-ray system shall comply with R9-7-419(B).Ariz. Admin. Code § R9-7-608
New Section R9-7-608 recodified from R12-1-608 at 24 A.A.R. 813, effective March 22, 2018 (Supp. 18-1).