Ariz. Admin. Code § 9-7-539

Current through Register Vol. 30, No. 45, November 8, 2024
Section R9-7-539 - Permanent Radiographic Installations
A. If a licensee maintains a permanent radiographic installation that does not fall within the definition of "enclosed radiography" in R9-7-102, the licensee shall ensure that each entrance, used for personnel access to the high radiation area, has either:
1. An entrance control device of the type described in R9-7-420(A)(1) that reduces the radiation level upon entry into the area, or
2. Both conspicuous visible and audible alarm signals to warn of the presence of radiation. The licensee shall ensure that the visible signal is actuated by radiation if a source is exposed and the audible signal is actuated if someone attempts to enter the installation while a source is exposed.
B. A licensee with an alarm signal shall test the alarm signal for proper operation with a radiation source each day before the installation is used for radiographic operations. The test shall include a check of both the visible and audible signals. A licensee with an entrance control device shall test the device monthly. If an entrance control device or alarm signal is operating improperly, the licensee shall immediately label the device or signal as "defective" and repair the device or signal within seven calendar days. The licensee may continue to use the facility during this seven-day period, if the licensee implements continuous surveillance requirements of R9-7-509 and uses an alarming rate meter.
C. A licensee shall maintain each record an alarm system or entrance control device test for three years after the record is made.

Ariz. Admin. Code § R9-7-539

New Section R9-7-539 recodified from R12-1-539 at 24 A.A.R. 813, effective March 22, 2018 (Supp. 18-1).