Ariz. Admin. Code § tit. 9, ch. 14, art. 6, tbl. 6.4

Current through Register Vol. 30, No. 45, November 8, 2024
Table 6.4 - Alternate Default Limits

QUALITY CONTROL PARAMETERS WITHOUT ACCEPTANCE CRITERIA SPECIFIED IN THE METHOD

DEFAULT LIMITS

Matrix Spike/LFM (processed or non-processed)

LCS/LFB

Matrix Spike/LCS for 8000 methods

±30%.

LCS/LFB (processed or non-processedVSecond source reference standard

CCV/continuing IPC

LOO/MRL (non-processed)

CCV/continuing IPC or ± 50%

LOO/MRL (processed)

LCS/LFB or ±50%

Methods that do not specify the LOO/MRL

± 50%

OCS (non-processed)

ICV/continuing IPC/manufacturer's limits

OCS (processed)

LCS/LFB/manufacturer's limits

IDOC limits

LFB/LCS

LFB/LCS/LFM/duplicate RPD

IDOC limits/?20%

Non-CCC compounds

CCC limits

ICV/CCV

± 10%

500, 600, 1600, and 8000 series methods that do not specify surrogates or acceptance limits for surrogates

70-130%.

500, 600, 1600, and 8000 series methods that do not specify internal standards or acceptance limits for internal standards

70-130%.

Methods that do not list a precision measurement

20% RPD

Ariz. Admin. Code tit. 9, ch. 14, art. 6, tbl. 6.4

Adopted by final rulemaking at 22 A.A.R. 2683, effective 10/1/2016.