Ariz. Admin. Code § tit. 13, ch. 10, art. 1, exh. H-4

Current through Register Vol. 30, No. 45, November 8, 2024
Exhibit H-4 - Standard Calibration Procedure Alco Sensor RBT AZ

THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ

CALIBRATION

AGENCY _____________________________________DATE __________________________

QA SPECIALIST ______________________________LOCATION ________________________

RBT AZ SERIAL # __________________________ALCO-SENSOR AZ SERIAL # ____________

() 1. Have a standard alcohol concentration solution ready.

This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0.______________ AC.

() 2. Depress RBT AZ ON button.

() 3. Depress Time button, enter PIN #, depress #1 button.

() 4. Follow RBT AZ and AS AZ display instructions.

() 5. Device temperature registers between 23° C and 27° C.

() 6. After a blank reading of 0.000 is displayed and the standard value is displayed, depress F3.

() 7. When AS AZ display flashes "CAL", introduce standard for 7 seconds; depress the MANUAL button on the AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)

() 8. Remove test record when printout is complete.

() 9. Run a calibration check on the Standard Calibration Check Procedure.

Test results: AC.

COMMENTS _____________________________________________________________________

_______________________________________________________________________________

_______________________________________________________________________________

_______________________________________________________________________________

SIGNATURE ____________________________________________________________________

DPS Form Exh H-4 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. H-4

New Exhibit H-4 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).