THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)
ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ
CALIBRATION
AGENCY _____________________________________DATE __________________________
QA SPECIALIST ______________________________LOCATION ________________________
RBT AZ SERIAL # __________________________ALCO-SENSOR AZ SERIAL # ____________
() 1. Have a standard alcohol concentration solution ready.
This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0.______________ AC.
() 2. Depress RBT AZ ON button.
() 3. Depress Time button, enter PIN #, depress #1 button.
() 4. Follow RBT AZ and AS AZ display instructions.
() 5. Device temperature registers between 23° C and 27° C.
() 6. After a blank reading of 0.000 is displayed and the standard value is displayed, depress F3.
() 7. When AS AZ display flashes "CAL", introduce standard for 7 seconds; depress the MANUAL button on the AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)
() 8. Remove test record when printout is complete.
() 9. Run a calibration check on the Standard Calibration Check Procedure.
Test results: AC.
COMMENTS _____________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
SIGNATURE ____________________________________________________________________
DPS Form Exh H-4 (Rev 05-01)
Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. H-4