Exhibit H-1 - Standard Operational Procedure Alco Sensor RBT AZ

OPERATIONAL CHECKLIST

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD OPERATIONAL PROCEDURE ALCO SENSOR RBT AZ

DUPLICATE BREATH TEST

SUBJECT NAME __________________________________DATE __________________________

AGENCY ___________________________________OPERATOR ___________________________

LOCATION _______________________________________________________________________

RBT AZ SERIAL # ____________________________ALCO SENSOR AZ SERIAL # ______________

TEST RESULTS 0. _________AC TIME _________

0. _________AC TIME _________

0. _________AC TIME _________

Immediately preceding administration of the tests, subject underwent at least a 15-minute deprivation period:

From

_______

to

_______

by

_______

(Time)

(Time)

(Name)

() 1. Depress RBT AZ ON button.

() 2. Depress zero set button, select subject or quick test.

() 3. Follow RBT AZ and AS AZ display instructions.

() 4. Enter case # &/or DL # if required.

() 5. Device temperature registers between 10° C and 40° C.

() 6.

a. If quick test, go to step 7.
b. If subject test, repeat steps 3 - 6 for duplicate test.
c. If the second subject test is not within 0.020 of the first test, repeat steps 3-6.
d. If the second subject test is within 0.020 of the first test, go to step 7.
e. If the third subject test, go to step 7.

() 7. Remove test record when printout is complete.

() 8. Turn off RBT AZ.

Note: Duplicate breath tests shall be administered at intervals of not less than 5 nor more than 10 minutes and the two consecutive tests shall agree within 0.020 alcohol concentration.

DPS Form Exh H-1 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. H-1

New Exhibit H-1 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).