Medical Devices; Therapeutic Devices; Classification of the Pediatric Continuous Renal Replacement Therapy System

Download PDF
Federal RegisterSep 16, 2024
89 Fed. Reg. 75493 (Sep. 16, 2024)
Document Headings

Document headings vary by document type but may contain the following:

  • the agency or agencies that issued and signed a document
  • the number of the CFR title and the number of each part the document amends, proposes to amend, or is directly related to
  • the agency docket number / agency internal file number
  • the RIN which identifies each regulatory action listed in the Unified Agenda of Federal Regulatory and Deregulatory Actions
  • See the Document Drafting Handbook for more details.

    Department of Health and Human Services Food and Drug Administration
  • 21 CFR Part 876
  • [Docket No. FDA-2024-N-4082]
  • AGENCY:

    Food and Drug Administration, HHS.

    ACTION:

    Final amendment; final order.

    SUMMARY:

    The Food and Drug Administration (FDA, Agency, or we) is classifying the pediatric continuous renal replacement therapy system into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the pediatric continuous renal replacement therapy system's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.

    DATES:

    This order is effective September 16, 2024. The classification was applicable on April 29, 2020.

    FOR FURTHER INFORMATION CONTACT:

    Gema Gonzalez, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 2530, Silver Spring, MD 20993-0002, 301-796-6519, Gema.Gonzalez@fda.hhs.gov.

    SUPPLEMENTARY INFORMATION:

    I. Background

    Upon request, FDA has classified the pediatric continuous renal replacement therapy system as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).

    FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)). Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the 510(k) process, when necessary, to market their device.

    II. De Novo Classification

    On October 9, 2018, FDA received Medtronic, Inc.'s request for De Novo classification of the CARPEDIEM System. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on April 29, 2020, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 876.5861. We have named the generic type of device pediatric continuous renal replacement therapy system, and it is identified as a device intended for use as an artificial kidney system for the management of pediatric patients with acute kidney injury and/or fluid overload by performing such therapies as hemodialysis, hemofiltration, hemodiafiltration, and isolated ultrafiltration. Using a hemodialyzer with a semipermeable membrane, the hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via ultrafiltration) and/or diffusion (via a concentration gradient in dialysate). The hemodialysis delivery machine, with an automated ultrafiltration controller, controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. During treatment, a patient's blood is circulated through the blood tubing set connected to the hemodialyzer's blood compartment. Blood access devices and accessories for hemodialysis required for the prescribed treatment are regulated under 21 CFR 876.5540.

    FDA notes that the ACTION caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1—Pediatric Continuous Renal Replacement Therapy System Risks and Mitigation Measures

    Identified risks to health Mitigation measures
    Adverse tissue reaction Biocompatibility evaluation, Pyrogenicity testing, and Non-clinical performance testing.
    Death Labeling, Clinical performance testing, and Usability testing.
    Infection Labeling, Reprocessing validation, Pyrogenicity testing, Shelf life testing, and Usability testing.
    Inadequate or incomplete treatment Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; Usability testing; and Software verification, validation, and hazard analysis.
    Clearance of essential blood substances or medications Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; Usability testing; and Software verification, validation, and hazard analysis.
    Blood loss or blood cell destruction Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; and Software verification, validation, and hazard analysis.
    Thermal injury Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; Usability testing; and Software verification, validation, and hazard analysis.
    Blood leak into the dialysis fluid Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; and Software verification, validation, and hazard analysis.
    Fluid imbalance Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; and Software verification, validation, and hazard analysis.
    Air embolism Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; Usability testing; and Software verification, validation, and hazard analysis.
    Fluid pump(s) reversal resulting in air infusion via the arterial bloodline Non-clinical performance testing; Clinical performance testing; Labeling; Shelf-life testing; Usability testing; and Software, verification, validation, and hazard analysis.
    Electrical shock Electrical safety testing.
    Electromagnetic interference with other devices/equipment Electromagnetic compatibility (EMC) testing.