Medical Devices; Reprocessed Single-Use Devices; Termination of Exemptions From Premarket Notification; Requirement for Submission of Validation Data

Download PDF
Federal RegisterApr 30, 2003
68 Fed. Reg. 23139 (Apr. 30, 2003)

AGENCY:

Food and Drug Administration, HHS.

ACTION:

Notice.

SUMMARY:

The Food and Drug Administration (FDA) is announcing the availability of a list (List I) of critical reprocessed single-use devices (SUDs) whose exemption from premarket submission is being terminated and for which validation data, as specified under the Medical Device User Fee and Modernization Act of 2002 (MDUFMA), is necessary in a premarket notification (510(k)); and a list (List II) of reprocessed SUDs that are currently subject to 510(k) requirements for which FDA has determined that validation data, as specified under MDUFMA, is necessary in a 510(k). FDA is requiring submission of these data to ensure that these reprocessed SUDs are substantially equivalent to predicate devices in accordance with MDUFMA.

DATES:

These actions are effective April 30, 2003. Manufacturers of SUDs identified in List I whose exemption is being terminated must submit 510(k)s for these devices by July 30, 2004, or their devices may no longer be marketed. Manufacturers who already have clearance letters for SUDs identified in List II must submit validation data for these devices by January 30, 2004, or marketing of these devices must cease.

ADDRESSES:

Submit written comments to the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. Submit electronic comments to http://www.fda.gov/dockets/ecomments . Comments on Lists I and II should be identified with the docket number found in brackets in the heading of this document.

FOR FURTHER INFORMATION CONTACT:

Barbara A. Zimmerman, Center for Devices and Radiological Health (HFZ-410), Food and Drug Administration, 9200 Corporate Blvd., Rockville, MD 20850.

SUPPLEMENTARY INFORMATION:

I. Background

On October 26, 2002, MDUFMA (Pub. L. 107-250), amended the Federal Food, Drug, and Cosmetic Act (the act) by adding section 510(o) (21 U.S.C. 360(o)), which provided new regulatory requirements for reprocessed SUDs. According to this new provision, in order to ensure that reprocessed SUDs are substantially equivalent to predicate devices, 510(k)s for certain reprocessed SUDs identified by FDA must include validation data. These required validation data include cleaning and sterilization data, and functional performance data demonstrating that each SUD will remain substantially equivalent to its predicate device after the maximum number of times the device is reprocessed as intended by the person submitting the premarket notification.

Before enactment of the new law, a manufacturer of a reprocessed SUD was required to obtain premarket approval or premarket clearance for the device, unless the device was exempt from premarket submission requirements. Under MDUFMA, some previously exempt reprocessed SUDs will no longer be exempt from premarket notification requirements. Manufacturers of these identified devices will need to submit 510(k)s that include validation data to be specified by FDA. Reprocessors of certain SUDs that are currently subject to cleared 510(k)s also will need to submit the validation data specified by the agency.

In the near future, FDA will publish a guidance document providing more specific information about the types of validation data that should be submitted in premarket notification submissions for the reprocessed SUDs listed in this notice.

A. Definitions

Under section 302(b) of MDUFMA, a reprocessed SUD is defined as an “original device that has previously been used on a patient and has been subjected to additional processing and manufacturing for the purpose of an additional single use on a patient. The subsequent processing and manufacture of a reprocessed single-use device shall result in a device that is reprocessed within the meaning of this definition.”

B. Reprocessed SUDs Exempt From Premarket Notification

Reprocessed SUDs are divided into three groups: (1) Critical, (2) semicritical, and (3) noncritical. The first two categories reflect definitions set forth in MDUFMA, and all three reflect a classification scheme recognized in the industry. These categories of devices are defined as follows:

Spaulding, E.H., “The Role of Chemical Disinfection in the Prevention of Nonsocomial Infections,” P.S. Brachman and T.C. Eickof (ed), Proceedings of International Conference on Nonsocomial Infections, 1970, American Hospital Association, Chicago, 1971:254-274.

(1) A critical reprocessed SUD is intended to contact normally sterile tissue or body spaces during use.

(2) A semicritical reprocessed SUD is intended to contact intact mucous membranes and not penetrate normally sterile areas of the body.

(3) A noncritical reprocessed SUD is intended to make topical contact and not penetrate intact skin.

1. Requirements for Critical Reprocessed SUDs

MDUFMA requires FDA to review the critical reprocessed SUDs that are currently exempt from premarket notification requirements and determine which of these devices require premarket notification to ensure their substantial equivalence to predicate devices. By April 26, 2003, FDA must identify in a Federal Register notice those critical reprocessed SUDs whose exemption from premarket notification will be terminated. List I in this Federal Register notice implements this MDUFMA requirement.

In accordance with MDUFMA, manufacturers of the devices identified in List I must submit 510(k)s that include validation data regarding cleaning, sterilization, and functional performance, in addition to all the other required elements of a 510(k) identified in § 807.87 (21 CFR 807.87), within 15 months of publication of this notice or no longer market their device.

2. Requirements for Semicritical Reprocessed SUDs

MDUFMA also requires FDA to review the semicritical reprocessed SUDs that are currently exempt from premarket notification requirements and determine which of these devices require premarket notification to ensure their substantial equivalence to predicate devices. FDA must identify these devices in a notice published in the Federal Register by April 26, 2004. Manufacturers of devices identified at that time will be required to submit 510(k)s that include validation data regarding cleaning, sterilization, and functional performance in addition to all the other required elements of a 510(k) identified in § 807.87, within 15 months of publication of that notice or no longer market their device.

3. Requirements for Noncritical Reprocessed SUDs

MDUFMA does not require FDA to take any action under this section for noncritical SUDs that are exempt from premarket submission requirements.

C. Reprocessed SUDs Already Subject to Premarket Notification Requirements

MDUFMA also requires FDA to review the types of reprocessed SUDs already subject to premarket notification requirements and to identify which of these devices require the submission of validation data to ensure their substantial equivalence to predicate devices. FDA must publish a list of these devices in the Federal Register by April 26, 2003, and update the list as necessary. List II of this Federal Register notice implements this MDUFMA requirement. The devices on List II may be critical, semicritical, or noncritical reprocessed SUDs.

1. For devices identified in List II that have not yet been cleared through the 510(k) process, manufacturers must submit 510(k)s that include validation data regarding cleaning, sterilization, and functional performance, in addition to all the other required elements of a 510(k) identified in § 807.87, upon publication of this notice in order to market these devices. FDA will soon publish guidance to help submitters understand what types of validation data should be included in these 510(k)s.

2. For devices identified in List II that already have been cleared through the 510(k) process, manufacturers must submit validation data regarding cleaning, sterilization, and functional performance within nine months of publication of this notice or marketing must cease. FDA will soon publish guidance to explain how a 510(k) holder may submit the additional data now being required to support an earlier clearance.

II. FDA's Implementation of New Section 510(o) of the Act

FDA used a number of criteria to determine which device types should be included in the lists required by MDUFMA. As part of its consideration, FDA relied upon the Review Prioritization Scheme (RPS) it described in the February 2000 draft guidance document entitled “Reprocessing and Reuse of Single-Use Devices: Review Prioritization Scheme.” In the RPS guidance, FDA set forth factors that could be used to evaluate risk associated with reprocessed SUDs. This approach assigned an overall risk to each SUD based on: (1) The risk of infection and (2) the risk of inadequate performance following reprocessing. Based on these risk factors, three categories of risk (high, moderate, and low) were developed. The designation of “high risk” was assigned to those devices that posed the greatest risk of infection and inadequate performance after reprocessing. In response to several comments about potential subjectivity of the RPS, FDA did not use the RPS approach when the agency finalized its enforcement priorities for reprocessed SUDs on August 14, 2000.

This draft guidance document is available on the CDRH Web site at http//www.fda.gov/cdrh/reuse/1156.pdf.

FDA has determined, however, that the RPS is an appropriate risk-based tool for developing the lists required by MDUFMA because the RPS identifies the devices that are likely to raise the most concerns about both infection transmission and inadequate performance following reprocessing. In formulating these lists, the agency also had the benefit of comments from stakeholders and an internal centerwide committee to evaluate the results of the RPS and ensure its consistency. In addition, there was a final review of all the devices on these lists by the Director of the Office of Device Evaluation. In this context, the agency believes these steps have adequately addressed concerns about the subjectivity of the RPS.

In addition to the previous criterion, FDA used one other criterion to identify those reprocessed SUDs that will be subject to the new requirements established by MDUFMA. The agency has included in these lists all reprocessed SUDs intended to come in contact with tissue at high risk of being infected with the causative agents of Creutzfeldt-Jakob Disease (CJD). These are generally devices intended for use in neurosurgery ophthalmology. This criterion was included in FDA's evaluation because insufficient scientific information exists at this time to establish standard methods to eliminate CJD infectious agents.

Therefore, in order to develop the two lists required by MDUFMA, FDA used the following process. First, the agency identified the types of SUDs that are being reprocessed. FDA did this by searching the 510(k) database for any 510(k)s that had been submitted for reprocessed SUDs and by asking original equipment manufacturers and reprocessors to provide information about types of devices that were being reprocessed. Second, FDA determined whether these devices are “critical,” “semi-critical,” or “non-critical”. (These definitions reflect the Spaulding classification and are the same definitions FDA used earlier in developing its RPS.) FDA then applied the criteria described previously and “listed” any reprocessed SUD that was either “high” risk according to the RPS or intended to come in contact with tissue at high risk of being infected with the causative agents of CJD.

Spaulding, E. H., “The Role of Chemical Disinfection in the Prevention of Nonsocomial Infections,” P. S. Brachman and T. C. Eickof (ed), Proceedings of International Conference on Nonsocomial Infections, 1970, American Hospital Association, Chicago, 1971:254-274.

All devices identified in List I (previously exempt from 510(k)) have been determined to be critical reprocessed SUDs. In addition to being critical, they are either high risk according to the RPS or intended to come in contact with tissue at high risk of being infected with CJD. It should be noted that not all exempt devices that are critical have been listed. Critical reprocessed SUDs that are not listed in List I at this time may be reconsidered in subsequent updates of the list. The devices in List II (devices currently subject to 510(k) requirements that now will require the submission of validation data) are either high risk according to the RPS or intended to come in contact with tissue at high risk of being infected with the causative agents of CJD.

FDA has also provided a reference list in Attachment 1. To show how FDA evaluated the risk of a specific device, Attachment 1 includes the entire group of devices FDA considered when identifying the reprocessed SUDs in Lists I and II, and shows how FDA applied the criteria that determined whether the device would be identified on either of these lists.

In the Federal Register of February 4, 2003 (68 FR 5643), FDA invited interested persons to provide information and share views on the implementation of MDUFMA. The agency received several comments that identified specific reprocessed SUDs to be included in Lists I and II. The agency considered these recommendations while finalizing this document. Although FDA's lists do not include all the reprocessed SUDs that were recommended, the agency believes that those devices that pose the greatest risk of infection transmission and inadequate performance have been identified. The agency recognizes, however, that these lists may need to be re-evaluated and updated over time. Therefore, FDA will consider comments from the public on additional devices that should be included in the lists at any time. The agency also notes that MDUFMA permits FDA to request validation data for a device type that is subject to 510(k) clearance but not yet included in List II. If this were to occur, FDA would ensure that manufacturers were aware of this change in the 510(k) submission requirements for that type of device by promptly updating the list.

Finally, FDA received one comment that suggested the agency's prior determinations about risk associated with reprocessed SUDs precluded FDA from now requiring 510(k)s for devices that were previously exempt or additional data for devices that were already cleared. FDA believes that this comment ignores the existence of MDUFMA's requirements. It is true that FDA had initially developed a regulatory approach for reprocessed SUDs that sought to treat those devices and original devices in a similar manner and that FDA had not required additional data to be submitted for certain reprocessed SUDs under that approach. However, through MDUFMA Congress clearly stated its intent to have the agency re-examine its policy with respect to reprocessed SUDs and legislated additional controls for those devices. FDA is committed to fulfilling its responsibilities under MDUFMA. The development and publication of these lists is part of the agency's implementation of these new statutory provisions.

List I.—Critical Reprocessed Single-Use Devices Previously Exempt From Premarket Notification Requirements That Will Now Require 510(k)s With Validation Data

[To be submitted by July 26, 2004]

21 CFR section Classification name Product code for Non-reprocessed device Product code for reprocessed device Product code name for reprocessed device
872.3240 Dental bur Diamond Coated NME Dental diamond coated bur.
872.4535 Dental diamond instrument DZP NLD Dental diamond instrument.
872.4730 Dental injection needle DZM NMW Dental needle.
874.4140 Ear, nose, and throat bur Microdebrider NLY ENT high speed microdebrider.
874.4140 Ear, nose, and throat bur Diamond Coated NLZ ENT diamond coated bur.
874.4420 Ear, nose, throat manual surgical * * * KAB, KBG, KCI NLB Laryngeal, Sinus, Tracheal trocar.
878.4200 Introduction/drainage catheter and accessories GCB NMT Catheter needle.
878.4800 Manual surgical instrument MJG NNA Percutaneous biopsy device.
878.4800 Manual surgical instrument FHR NMU Gastro-Urology needle.
878.4800 Manual surgical instrument for * * * DWO NLK Cardiovascular biopsy needle.
878.4800 Manual surgical instrument for * * * GAA NNC Aspiration and injection.
882.4190 Forming/cutting clip instrument HBS NMN Forming/cutting clip instrument.
884.1730 Laparoscopic insufflator * * * HIF NMI Laparoscopic insufflator and accessories.
884.4530 OB/GYN specialized manual instrument HFB NMG Gynecological biopsy forceps.
886.4350 Manual ophthalmic surgical instrument HNN NLA Ophthalmic knife.

List II.—Reprocessed Single-Use Devices Subject to Premarket Notification Requirements That Will Now Require The Submission of Validation Data

[Manufacturers who already have 510(k) clearance for these devices must submit validation data by January 26, 2004. Any new 510(k) will require validation data upon publication of this list.]

21 CFR section Classification name Product code for non-reprocessed device Product code for reprocessed device Product code name for reprocessed device
Unclassified Oocyte aspiration needles MHK NMO Oocyte aspiration needles.
Unclassified Percutaneous transluminal angioplasty catheter angioplasty catheter LIT NMM Transluminal peripheral angioplasty catheter.
Unclassified Ultrasonic surgical instrument LFL NLQ Ultrasonic scalpel.
868.5150 Anesthesia conduction needle BSP NNH Anesthetic conduction needle (with or without introducer).
868.5150 Anesthesia conduction needle MIA NMR Short term spinal needle.
868.5730 Tracheal tube BTR NMA Tracheal tube (with or without connector).
868.5905 Non­con­tin­u­ous (IPPB) BZD NMC Noncontinuous ventilator (respirator) mask.
870.1200 Diagnostic intravascular catheter DQO NLI Angiography catheter.
870.1220 Electrode recording catheter DRF NLH Electrode recording catheter.
870.1220 Electrode recording catheter MTD NLG Intracardiac mapping catheter.
870.1230 Fiberoptic oximeter catheter DQE NMB Fiberoptic oximeter catheter.
870.1280 Steerable catheter DRA NKS Steerable catheter.
870.1290 Steerable catheter control system DXX NKR Steerable catheter control system.
870.1330 Catheter guide wire DQX NKQ Catheter guide wire.
870.1390 Trocar DRC NMK Cardiovascular trocar.
870.1650 Angiographic injector and syringe DXT NKT Angiographic injector and syringe.
870.1670 Syringe actuator for injector DQF NKW Injector for actuator syringe.
870.2700 Oximeter MUD NMD Tissue saturation oximeter.
870.2700 Oximeter DQA NLF Oximeter.
870.3535 Intra-aortic balloon and control system DSP NKO Intra-aortic balloon and control system.
870.4450 Vascular clamp DXC NMF Vascular clamp.
870.4885 External vein stripper DWQ NLJ External vein stripper.
872.5470 Orthodontic plastic bracket DYW NLC Orthodontic plastic bracket.
874.4680 Bronchoscope (flexible or rigid) and accessories BWH NLE Bronchoscope (nonrigid) biopsy forceps.
876.1075 Gastro-Urology biopsy instrument FCG NMX G-U biopsy needle and needle set.
876.1075 Gastroenterology-urology biopsy instrument KNW NLS Biopsy instrument.
876.1500 Endoscope and accessories FBK, FHP NMY Endoscopic needle.
876.1500 Endoscope and accessories MPA NKZ Endoilluminator.
876.1500 Endoscope and accessories GCJ NLM General and plastic surgery laparoscope.
876.1500 Endoscope and accessories FHO NLX Spring-loaded Pneumoperitoneum Needle.
876.4300 Endoscopic electrosurgical unit and accessories FAS NLW Active urological electrosurgical electrode.
876.4300 Endoscopic unit accessories FEH NLV Flexible suction coagulator electrode.
876.4300 Endoscopic electrosurgical unit and accessories KGE NLU Electric biopsy forceps.
876.4300 Endoscopic electrosurgical unit and accessories FDI NLT Flexible snare.
876.4300 Endoscopic electrosurgical unit and accessories KNS NLR Endoscopic (with or without accessories) Electrosurgical unit.
876.5010 Biliary catheter and accessories FGE NML Biliary catheter.
876.5540 Blood access device and accessories LBW NNF Single needle dialysis set (co-axial flow).
876.5540 Blood access device and accessories FIE NNE Fistula needle.
876.5820 Hemodialysis systems and accessories FIF NNG Single needle dialysis set with and accessories uni-directional pump.
878.4300 Implantable clip FZP NMJ Implantable clip.
878.4750 Implantable staple GDW NLL Implantable staple.
880.5570 Hypodermic single lumen needle FMI NKK Hypodermic single lumen needle.
880.5860 Piston syringe FMF NKN Piston syringe.
882.4300 Manual cranial drills, burrs, trephines and accessories HBG NLO (Manual) drills, burrs, burrs, trephines and accessories.
882.4305 Powered compound cranial drills, burrs, trephines . . . HBF NLP (Powered, compound) drills, burrs, trephines and accessories.
882.4310 Powered simple cranial drills, burrs, trephines HBE NLN (Simple, powered) drills, burrs, trephines and accessories.
884.1720 Gynecologic laparoscope and accessories HET NMH Gynecologic laparoscope (and accessories).
884.6100 Assisted reproduction needle MQE NNB Assisted reproduction needle.
886.4370 Keratome HMY, HNO NKY Keratome blade.
886.4670 Phacofragmentation system HQC NKX Phacoemulsification needle.
892.5730 Radionuclide brachytherapy source IWF NMP Isotope needle.
Hemodialyzers have been excluded from this list because the reuse of hemodialyzers is addressed in “Guidance for Hemodialyzer Reuse Labeling” (final draft issued on October 6, 1995).

III. Comments

You may submit written or electronic comments on these lists to the Dockets Management Branch (see ADDRESSES). You may submit a single copy of an electronic comment to http://www.fda.gov/dockets/ecomments . You should submit two copies of any mailed comments but individuals may submit one copy. You should identify your comment with the docket number found in brackets in the heading of this document. You may see any comments FDA receives in the Dockets Management Branch between 9 a.m. and 4 p.m., Monday through Friday.

Attachment 1.—List of SUDs Known to be Reprocessed or Considered for Reprocessing

2121
Medical specialty Device type Regulation No. Class Product code Risk Critical/semi-critical/non-critical Premarket exempt
1 Cardio Cardiopulmonary Bypass Marker Unclassified MAB 1 C N
2 Cardio Percutaneous & Operative Transluminal Coronary Angioplasty Catheter (PCTA) post amendment III LOX 3 C N
3 Cardio Percutaneous Ablation Electrode Post amendment III LPB 3 C N
4 Cardio Peripheral Transluminal Angioplasty (PTA) Catheter Unclassified LIT 3 C N
5 Cardio Blood-Pressure Cuff 870.1120 II DXQ 1 N N
6 Cardio Angiography Catheter 870.1200 II DQO 3 C N
7 Cardio Electrode Recording Catheter 870.1220 II DRF 3 C N
8 Cardio High-Density Array Catheter 870.1220 II MTD 3 C N
9 Cardio Fiberoptic Oximeter Catheter 870.1230 II DQE 3 C N
10 Cardio Steerable Catheter 870.1280 II DRA 3 C N
11 Cardio Steerable Catheter Control System 870.1290 II DXX 3 C N
12 Cardio Guide Wire 870.1330 II DQX 3 C N
13 Cardio Angiographic Needle 870.1390 II DRC 3 C N
14 Cardio Trocar 870.1390 II DRC 3 C N
15 Cardio Syringes 870.1650 II DXT 3 C N
16 Cardio Injector Type Syringe Actuator 870.1670 II DQF 3 C N
17 Cardio Oximeter 870.2700 II DQA 3 N N
18 Cardio Tissue Saturation Oximeter 870.2700 II MUD 3 C N
19 Cardio Intra-Aortic Balloon System 870.3535 III DSP 3 C N
20 Cardio Vascular Clamp 870.4450 II DXC 3 C N
21 Cardio Device, Stabilizer, Heart 870.4500 I MWS 2 C Y
22 Cardio External Vein Stripper 870.4885 II DWQ 3 C N
23 Cardio Compressible Limb Sleeve 870.5800 II JOW 1 N N
24 Dental Bur 872.3240 I EJL 1 C Y
25 Dental Diamond Coated Bur 872.3240 I EJL 3 C Y
26 Dental Diamond Instrument 872.4535 I DZP 3 C Y
27 Dental AC-Powered Bone Saw 872.4120 II DZH 2 C N
28 Dental Manual Bone Drill and Wire Driver 872.4120 II DZJ 2 C N
29 Dental Powered Bone Drill 872.4120 II DZI 2 C N
30 Dental Intraoral Drill 872.4130 I DZA 1 C Y
31 Dental Injection Needle 872.4730 I DZM 3 C Y
32 Dental Metal Orthodontic Bracket 872.5410 I EJF 3 S Y
33 Dental Plastic Orthodontic Bracket 872.5470 II DYW 3 S N
34 ENT Bur 874.4140 I EQJ 1 C Y
35 ENT Diamond Coated Bur 874.4140 I EQJ 3 C Y
36 ENT Microdebrider 874.4140 I EQJ 3 C Y
37 ENT Microsurgical Argon Fiber Optic Laser Cable, For Uses Other Than Otology, Including Laryngology & General Use In Otolarngology 874.4490 II LMS 1 S N
38 ENT Microsurgical Argon Fiber Optic Laser Cable, For Use In Otology 874.4490 II LXR 1 S N
39 ENT Microsurgical Carbon-Dioxide Fiber Optic Laser Cable 874.4500 II EWG 1 S N
40 ENT Bronchoscope Biopsy Forceps (Non-Rigid) 874.4680 II BWH 3 S N
41 ENT Bronchoscope Biopsy Forceps (Rigid) 874.4680 II JEK 1 S N
42 Gastro/Urology Biopsy Forceps Cover 876.1075 I FFF 1 S Y
43 Gastro/Urology Biopsy Instrument 876.1075 II KNW 3 S N
44 Gastro/Urology Biopsy Needle Set 876.1075 II FCG 3 S N
45 Gastro/Urology Biopsy Punch 876.1075 II FCI 2 S N
46 Gastro/Urology Mechanical Biopsy Instrument 876.1075 II FCF 2 S N
47 Gastro/Urology Non-Electric Biopsy Forceps 876.1075 I FCL 3 S Y
48 Gastro/Urology Cytology Brush For Endoscope 876.1500 II FDX 2 S N
49 Gastro/Urology Endoscope Accessories 876.1500 II KOG 2 S N
50 Gastro/Urology Extraction Balloons/Baskets 876.1500 II KOG 2 S N
51 Gastro/Urology Endoscopic Needle 876.1500 II FBK 3 C N
52 Gastro/Urology Simple Pneumoperitoneum Needle 876.1500 II FHP 3 C N
53 Gastro/Urology Spring Loaded Pneumoperitoneum Needle 876.1500 II FHO 3 C N
54 Gastro/Urology Active Electrosurgical Electrode 876.4300 II FAS 3 S N
55 Gastro/Urology Biliary Sphincterotomes 876.5010, 876.1500 II FGE 3 S N
56 Gastro/Urology Electric Biopsy Forceps 876.4300 II KGE 3 S N
57 Gastro/Urology Electrosurgical Endoscopic Unit (With Or Without Accessories) 876.4300 II KNS 3 S N
58 Gastro/Urology Flexible Snare 876.4300 II FDI 3 S N
59 Gastro/Urology Flexible Suction Coagulator Electrode 876.4300 II FEH 3 S N
60 Gastro/Urology Flexible Stone Dislodger 876.4680 II FGO 3 S Y
61 Gastro/Urology Metal Stone Dislodger 876.4680 II FFL 3 S Y
62 Gastro/Urology Needle Holder 876.4730 I FHQ 1 C Y
63 Gastro/Urology Non-Electrical Snare 876.4730 I FGX 1 S Y
64 Gastro/Urology Urological Catheter 876.5130 II KOD 2 S N
65 Gastro/Urology Single Needle Dialysis Set 876.5540 II LBW, FIE 3 C N
66 Gastro/Urology Hemodialysis Blood Circuit Accessories 876.5820 II KOC 2 S N
67 Gastro/Urology Single Needle Dialysis Set 876.5820 II FIF 3 C N
68 GE/U Hemorrhoidal Ligator 876.4400 II FHN 2 C N
69 General Hospital Implanted, Programmable Infusion Pump Post-amendment III LKK 3 C N
70 General Hospital Needle Destruction Device Post-amendment III MTV 1 N N
71 General Hospital Non-Powered Flotation Therapy Mattress 880.5150 I IKY 2 N Y
72 General Hospital Non AC-Powered Patient Lift 880.5510 I FSA N Y
73 General Hospital Alternating Pressure Air Flotation Mattress 880.5550 II FNM 1 N Y
74 General Hospital Temperature Regulated Water Mattress 880.5560 I FOH 2 N Y
75 General Hospital Hypodermic Single Lumen Needle 880.5570 II FMI 3 C N
76 General Hospital Piston Syringe 880.5860 II FMF 3 C N
77 General Hospital Mattress Cover (Medical Purposes) 880.6190 I FMW 2 N Y
78 General Hospital Disposable Medical Scissors 880.6820 I JOK N Y
79 General Hospital Irrigating Syringe 880.6960 I KYZ, KYY 1 C Y
80 Infection Control Surgical Gowns 878.4040 II FYA 1 C N
81 Lab Blood Lancet 878.4800 I FMK 1 C Y
82 Neuro Clip Forming/Cutting Instrument, 882.4190 I HBS 3* C Y
83 Neuro Drills, Burrs, Trephines & Accessories (Manual) 882.4300 II HBG 3* C N
84 Neuro Drills, Burrs, Trephines & Accessories (Compound, Powered) 882.4305 II HBF 3* C N
85 Neuro Drills, Burrs, Trephines & Accessories (Simple, Powered) 882.4310 II HBE 3* C N
86 OB/GYN Oocyte aspiration needle Unclassified II MHK 3 C N
87 OB/GYN Laparoscope Accessories 884.1720 I HET 2 C Y
88 OB/GYN Laparoscope Accessories 884.1720 II HET 3 C N
89 OB/GYN Laparoscopic Dissectors 884.1720 I HET 2 C Y
90 OB/GYN Laparoscopic Graspers 884.1720 I HET 2 C Y
91 OB/GYN Laparoscopic Scissors 884.1720 I HET 2 C Y
92 OB/GYN Insufflator Accessories (tubing, Verres needle, kits) 884.1730 II HIF 3 C Y
93 OB/GYN Laparoscopic Insufflator 884.1730 II HIF 2 N N
94 OB/GYN Endoscopic Electrocautery And Accessories 884.4100 II HIM 2 N N
95 OB/GYN Gynecologic Electrocautery (And Accessories) 884.4120 II HGI 2 N N
96 OB/GYN Endoscopic Bipolar Coagulator-Cutter (And Accessories) 884.4150 II HIN 2 N N
97 OB/GYN Culdoscopic Coagulator (And Accessories) 884.4160 II HFI 2 N N
98 OB/GYN Endoscopic Unipolar Coagulator-Cutter (And Accessories) 884.4160 II KNF 2 N N
99 OB/GYN Hysteroscopic Coagulator (And Accessories) 884.4160 II HFH 2 N N
100 OB/GYN Unipolar Laparoscopic Coagulator (And Accessories) 884.4160 II HFG 2 N N
101 OB/GYN Episiotomy Scissors 884.4520 I HDK 1 C Y
102 OB/GYN Umbilical Scissors 884.4520 I HDJ 1 C Y
103 OB/GYN Biopsy Forceps 884.4530 I HFB 3 C Y
104 OB/GYN Assisted reproduction needle 884.6100 II MQE 3 C N
105 Ophthalmic Endoilluminator 876.1500 II MPA 3* C N
106 Ophthalmic Surgical Drapes 878.4370 II KKX 2 C N
107 Ophthalmic Ophthalmic Knife 886.4350 I HNN 3 C Y
108 Ophthalmic Keratome Blade 886.4370 Inot exempt HMY, HNO 3 C N
109 Ophthalmic Phacoemulsification Needle 886.4670 II HQC 3 C N
110 Ophthalmic Phacoemulsification/Phacofragmentation Fluidic 886.4670 II MUS 2 C N
111 Ophthalmic Phacofragmentation Unit 886.4670 II HQC 1 N N
112 Ortho Saw Blades 878.4820 I GFA, DWH, GEY, GET 1 C Y
113 Ortho Surgical Drills 878.4820 I GEY, GET 1 C Y
114 Ortho Arthroscope accessories 888.1100 II HRX 2 C Y
115 Ortho Bone Tap 888.4540 I HWX 1 C Y
116 Ortho Burr 888.4540 I HTT 1 C Y
117 Ortho Carpal Tunnel Blade 888.4540 I LXH 2 C Y
118 Ortho Countersink 888.4540 I HWW 1 C Y
119 Ortho Drill Bit 888.4540 I HTW 1 C Y
120 Ortho Knife 888.4540 I HTS 1 C Y
121 Ortho Manual Surgical Instrument 888.4540 I LXH 1 C Y
122 Ortho Needle Holder 888.4540 I HXK 1 C Y
123 Ortho Reamer 888.4540 I HTO 1 C Y
124 Ortho Rongeur 888.4540 I HTX 1 C Y
125 Ortho Scissors 888.4540 I HRR 1 C Y
126 Ortho Staple Driver 888.4540 I HXJ 1 C Y
127 Ortho Trephine 888.4540 I HWK 1 C Y
128 Ortho Flexible Reamers/Drills 886.4070 878.4820 I GEY, HRG 1 C Y
129 Ortho External Fixation Frame 888.3040 888.3030 II JEC KTW KTT 2 N N
130 Physical Medicine Non-Heating Lamp for Adjunctive Use Inpatient Therapy Unclassified NHN 1 N N
131 Physical Medicine Electrode Cable, 890.1175 II IKD 1 N Y
132 Physical Medicine External Limb Component, Hip Joint 890.3420 I ISL 2 N Y
133 Physical Medicine External Limb Component, Knee Joint 890.3420 I ISY N Y
134 Physical Medicine External Limb Component, Mechanical Wrist 890.3420 I ISZ 2 N Y
135 Physical Medicine External Limb Component, Shoulder Joint 890.3420 I IQQ 2 N Y
136 Plastic Surgery Stapler 878.4800 I GAG, GEF, FHM, HBT 2 C Y
137 Radiology Isotope Needle 892.5730 II IWF 3 C N
138 Resp Endotracheal Tube Changer Unclassified III LNZ 3 C N
139 Resp Anesthesia conduction needle 868.5150 II BSP 3 C N
140 Resp Short term spinal needle 868.5150 II MIA 3 C N
141 Resp Respiratory Therapy And Anesthesia Breathing Circuits 868.5240 I CAI 2 S Y
142 Resp Oral And Nasal Catheters 868.5350 I BZB 1 C Y
143 Resp Gas Masks 868.5550 I BSJ 1 S Y
144 Resp Breathing Mouthpiece 868.5620 I BYP 1 N Y
145 Resp Tracheal Tube 868.5730 II BTR 3 C N
146 Resp Airway Connector 868.5810 I BZA 2 S Y
147 Resp CPAP Mask 868.5905 II BZD 3 S N
148 Resp Emergency Manual Resuscitator 868.5915 II BTM 2 S N
149 Resp Tracheobronchial Suction Catheter 868.6810 I BSY 3 S Y
150 Surgery AC-powered Orthopedic Instrument and accessories Unclassified HWE 2 C N
151 Surgery Breast Implant Mammary Sizer Unclassified MRD 1 C N
152 Surgery Ultrasonic Surgical Instrument Unclassified LFL 3 C N
153 Surgery Trocar 874.4420 I KAB, KBG, KCI 3 C Y
154 Surgery Endoscopic Blades 876.1500 II GCP, GCR 2 C N
155 Surgery Endoscopic Guidewires 876.1500 II GCP, GCR 1 C N
156 Surgery Inflatable External Extremity Splint 878.3900 I FZF 1 N Y
157 Surgery Noninflatable External Extremity Splint 878.3910 I FYH 1 N Y
158 Surgery Catheter needle 878.4200 I GCB 3 C Y
159 Surgery Implantable Clip 878.4300 II FZP 3 C N
160 Surgery Electrosurgical And Coagulation Unit With Accessories 878.4400 II BWA 2 C N
161 Surgery Electrosurgical Apparatus 878.4400 II HAM 2 C N
162 Surgery Electrosurgical Cutting & Coagulation Device & Accessories 878.4400 II GEI 2 C N
163 Surgery Electrosurgical Device 878.4400 II DWG 2 C N
164 Surgery Electrosurgical Electrode 878.4400 II JOS 2 C N
165 Surgery Implantable Staple, Clamp, Clip for Suturing Apparatus 878.4750 II GDW 3 C N
166 Surgery Percutaneous biopsy device 878.4800 I MJG 3 C Y
167 Surgery Gastro-Urology needle 878.4800 I FHR 3 C Y
168 Surgery Aspiration and injection needle 878.4800 I GAA 3 C Y
169 Surgery Biopsy Brush 878.4800 I GEE 1 C Y
170 Surgery Blood Lancet 878.4800 I FMK 1 C Y
171 Surgery Bone Hook 878.4800 I KIK 1 C Y
172 Surgery Cardiovascular Biopsy Needle 878.4800 I DWO 3 C Y
173 Surgery Clamp 878.4800 I GDJ 1 C Y
174 Surgery Clamp 878.4800 I HXD 1 C Y
175 Surgery Curette 878.4800 I HTF 1 C Y
176 Surgery Disposable Surgical Instrument 878.4800 I KDC 1 C Y
177 Surgery Disposable Vein Stripper 878.4800 I GAJ 1 C Y
178 Surgery Dissector 878.4800 I GDI 1 C Y
179 Surgery Forceps 878.4800 I GEN 2 C Y
180 Surgery Forceps 878.4800 I HTD 2 C Y
181 Surgery Gouge 878.4800 I GDH 1 C Y
182 Surgery Hemostatic Clip Applier 878.4800 I HBT 2 C Y
183 Surgery Hook 878.4800 I GDG 1 C Y
184 Surgery Manual Instrument 878.4800 I MDM, MDW 1 C Y
185 Surgery Manual Retractor 878.4800 I GZW 1 C Y
186 Surgery Manual Saw And Accessories 878.4800 I GDR HAC 1 C Y
187 Surgery Manual Saw And Accessories 878.4800 I HAC 1 C Y
188 Surgery Manual Surgical Chisel 878.4800 I FZO 1 C Y
189 Surgery Mastoid Chisel 878.4800 I JYD 1 C Y
190 Surgery Orthopedic Cutting Instrument 878.4800 I HTZ 1 C Y
191 Surgery Orthopedic Spatula 878.4800 I HXR 1 C Y
192 Surgery Osteotome 878.4800 I HWM 1 C Y
193 Surgery Rasp 878.4800 I GAC 1 C Y
194 Surgery Rasp 878.4800 I HTR 1 C Y
195 Surgery Retractor 878.4800 I GAD 1 C Y
196 Surgery Retractor 878.4800 I HXM 1 C Y
197 Surgery Saw 878.4800 I HSO 1 C Y
198 Surgery Scalpel Blade 878.4800 I GES 1 C Y
199 Surgery Scalpel Handle 878.4800 I GDZ 1 C Y
200 Surgery Scissors 878.4800 I LRW 1 C Y
201 Surgery Snare 878.4800 I GAE 1 C Y
202 Surgery Spatula 878.4800 I GAF 1 C Y
203 Surgery Staple Applier 878.4800 I GEF 2 C Y
204 Surgery Stapler 878.4800 I GAG 2 C Y
205 Surgery Stomach And Intestinal Suturing Apparatus 878.4800 I FHM 2 C Y
206 Surgery Surgical Curette 878.4800 I FZS 1 C Y
207 Surgery Surgical Cutter 878.4800 I FZT 1 C Y
208 Surgery Surgical Knife 878.4800 I EMF 1 S Y
209 Surgery Laser Powered Instrument 878.4810 II GEX 2 C N
210 Surgery Ac-Powered Motor 878.4820 I GEY 2 C Y
211 Surgery Bit 878.4820 I GFG 1 C Y
212 Surgery Bur 878.4820 I GFF, GEY 1 C Y
213 Surgery Cardiovascular Surgical Saw Blade 878.4820 I DWH 1 C Y
214 Surgery Chisel (Osteotome) 878.4820 I KDG 1 C Y
215 Surgery Dermatome 878.4820 I GFD 1 C Y
216 Surgery Electrically Powered Saw 878.4820 I DWI 2 C Y
217 Surgery Pneumatic Powered Motor 878.4820 I GET 2 C Y
218 Surgery Pneumatically Powered Saw 878.4820 I KFK 2 C Y
219 Surgery Powered Saw And Accessories 878.4820 I HAB 2 C Y
220 Surgery Saw Blade 878.4820 I GFA 1 C Y
221 Surgery Nonpneumatic Tourniquet 878.5900 I GAX 1 N Y
222 Surgery Pneumatic Tourniquet 878.5910 I KCY N Y
223 Surgery Enodoscopic Staplers 888.4540 I HXJ 2 C Y
224 Surgery Trocar 876.1500 870.1390 II GCJ, DRC 3 C N
225 Surgery Surgical Cutting Accessories 878.4800, 874.4420 I GDZ, GDX, GES, KBQ, KAS 2 C Y
226 Surgery Electrosurgical Electrodes/Handles/Pencils 876.4300 878.4400 II HAM, GEI, FAS 2 C N
227 Surgery Scissor Tips 878.4800, 884.4520, 874.4420 I LRW, HDK, HDJ, JZB, KBD 2 C Y
228 Surgery Laser Fiber Delivery Systems 878.4810 874.4500 886.4390 884.4550 886.4690 II GEX EWG LLW HQF HHR HQB 1 C N
1 = low risk according to RPS
2 = moderate risk according to RPS
3 = high risk according to RPS
3* = high risk due to neurological use

Dated: April 23, 2003.

Jeffrey Shuren,

Assistant Commissioner for Policy.

[FR Doc. 03-10413 Filed 4-23-03; 5:03 pm]

BILLING CODE 4160-01-P