AGENCY:
Food and Drug Administration, HHS.
ACTION:
Final amendment; final order.
SUMMARY:
The Food and Drug Administration (FDA or we) is classifying the hemodialyzer with expanded solute removal profile into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the hemodialyzer with expanded solute removal profile's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.
DATES:
This order is effective September 6, 2024. The classification was applicable on August 28, 2020.
FOR FURTHER INFORMATION CONTACT:
Jade Noble, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 2684, Silver Spring, MD 20993-0002, 240-402-5077, Jade.Noble@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the hemodialyzer with expanded solute removal profile as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness.
The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)). Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.
When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the 510(k) process, when necessary, to market their device.
II. De Novo Classification
On September 16, 2019, FDA received Baxter Healthcare Corporation's request for De Novo classification of the Theranova Dialyzers (Theranova 400, Theranova 500). FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.
Therefore, on August 28, 2020, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 876.5862. We have named the generic type of device hemodialyzer with expanded solute removal profile, and it is identified as a device intended for use as part of an artificial kidney system for the treatment of patient with renal failure by performing such therapies as hemodialysis, hemofiltration, and hemodiafiltration. A hemodialyzer with expanded solute removal profile includes modifications to the semipermeable membrane that allows for increased removal or uremic retention solutes compared with standard high-flux hemodialyzers of the high permeability hemodialysis system classification (§ 876.5860 (21 CFR 876.5860)) including solutes at the upper end of the “middle” molecular weight range (0.5 kDa to 60 kDa). This device is intended to be used with the extracorporeal hemodialysis delivery systems, blood tubing sets, blood access devices, and accessories regulated under 21 CFR 876.5820, 876.5860, 876.5540, and/or 876.5600.
FDA notes that the ACTION caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” Beginning in December 2019, this editorial change was made to indicate that the document “amends” the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.
FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.
Table 1—Hemodialyzer With Expanded Solute Removal Profile Risks and Mitigation Measures
Identified risks to health | Mitigation measures |
---|---|
Adverse tissue reaction | Biocompatibility evaluation, Pyrogenicity testing, and Non-clinical performance testing. |
Infection or pyrogen reaction | Labeling, Pyrogenicity testing, Sterilization validation, Non-clinical performance testing, and Shelf-life testing. |
Inadequate or incomplete treatment | Non-clinical performance testing, Labeling, and Shelf-life testing. |
Clearance of essential blood substances or medications | Non-clinical performance testing, Clinical performance testing, Labeling, and Shelf-life testing. |
Blood loss or blood cell destruction | Non-clinical performance testing, Labeling, and Shelf-life testing. |
Blood leak into the dialysis fluid | Non-clinical performance testing, Labeling, and Shelf-life testing. |
Air or particle embolism | Non-clinical performance testing, Labeling, and Shelf-life testing. |
Fluid imbalance | Non-clinical performance testing and Labeling. |
Acid-base imbalance | Non-clinical performance testing and Labeling. |