Ex Parte UkawaDownload PDFPatent Trial and Appeal BoardApr 12, 201612638658 (P.T.A.B. Apr. 12, 2016) Copy Citation UNITED STA TES p A TENT AND TRADEMARK OFFICE APPLICATION NO. FILING DATE FIRST NAMED INVENTOR 12/638,658 12/15/2009 TeijiUKAWA 39083 7590 04/14/2016 KENEALY VAIDYA LLP 3000 K Street, N.W. Suite 310 Washington, DC 20007 UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www .uspto.gov ATTORNEY DOCKET NO. CONFIRMATION NO. 6002-0041 5439 EXAMINER BURRAGE, MICHAEL J ART UNIT PAPER NUMBER WMB NOTIFICATION DATE DELIVERY MODE 04/14/2016 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address( es): avaidya@kviplaw.com uspto@kviplaw.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD Ex parte TEIJI UKA WA Appeal2014-000611 Application 12/638,658 Technology Center 3700 Before ERIC B. GRIMES, ULRIKE W. JENKS, and ROBERT A. POLLOCK, Administrative Patent Judges. PERCURIAM. DECISION ON APPEAL This is a decision on appeal 1 under 35 U.S.C. § 134(a) from the Examiner's rejection of claims 1-15. We have jurisdiction under 35 U.S.C. § 6(b ). We AFFIRM. STATEMENT OF THE CASE The Specification discloses "a respiratory function measuring apparatus which can measure a respiratory function signal indicative of an intrathoracic pressure change and the like from the invasive blood pressure" 1 Appellant identifies the Real Party in Interest as Nihon Kohden Corporation (App. Br. 3). Appeal2014-000611 Application 12/638,658 (Spec. 1 i-f 1).2 "In the measurement of work of breathing, the measurement of the intrathoracic pressure is necessary, but it is difficult to measure the intrathoracic pressure .... In the vicinity of the right atrium, particularly, the central venous pressure is low, and hence strongly reflects the intrathoracic pressure" (id. at 1 i-fi-12 and 3). The Specification discloses "a respiratory function measuring apparatus in which components derived from cardiac contraction can be removed from a central venous pressure waveform ... whereby the respiratory variation of the intrathoracic pressure can be easily estimated" (id. at 2 i-f 6). Claim 1, the only independent claim, is representative of the claims on appeal and reads as follows: 1. A respiratory function measuring apparatus comprising: a first sensor configured to detect an invasive blood pressure; a second sensor configured to measure frequency of at least one of a heart beat and a respiration; and a controller configured to obtain a respiratory function signal from the invasive blood pressure detected by the first sensor, by using at least one of the frequency measured by the second sensor and a harmonic of the frequency. 2 References to the Specification herein refer to the substitute Specification filed on July 12, 2012. 2 Appeal2014-000611 Application 12/638,658 Issue The claims stand rejected under 35 U.S.C. § 103(a) as foIIows: 1. Claims 1-10 in view of Danehorn3 and Liao4 ; IL Claim 11 in view ofDanehorn, Liao, and Joeken5; !fl. Claims 12 and 15 in view of Danehom, Liao, and Euliano6; and IV Claims 13 and 14 in view ofDanehom, Liao, and Anderson.7 I. The Examiner has rejected claims 1-10 under 35 U.S.C. § 103(a) as obvious over Danehom and Liao (Ans. 3---6). The issue presented is: Does the evidence of record support the Examiner's conclusion that the combination of Danehom and Liao would have made obvious a respiratory function measuring apparatus having "a controller configured to obtain a respiratory function signal from the invasive blood pressure detected by the first sensor, by using at least one of the frequency measured by the second sensor and a harmonic of the frequency?" 3 Kenneth Danehom et al., US 2007/0073170 Al, published Mar. 29, 2007 ("Dane horn"). 4 Wangcai Liao et al., US 2008/0243016 Al, published Oct. 2, 2008 ("Liao"). 5 Stephan Joeken, US 2008/0033306 Al, published Feb. 7, 2008. 6 Neil R. Euliano et al., US 2004/0040560 Al, published Mar. 4, 2004 ("Euliano"). 7 Catherine A. Anderson et al., US 4,930,519, issued June 5, 1990 ("Anderson"). 3 Appeal2014-000611 Application 12/638,658 Findings of Fact We adopt the Examiner's findings and analysis concerning the scope and content of the prior art. The following facts are repeated for reference convemence. 1. Danehom teaches a "device for removing artefacts caused by patient respiration in measured blood pressure data and in particular in blood pressure data acquired invasively in the heart and/or an artery of the patient" (Danehom i-f 2; see Ans. 3--4). 2. Danehom teaches that the level of C02 in the patient's expired air is acquired as a respiratory signal during the blood pressure measurement, and the respiratory signal is used to approximate and remove the artefacts caused by patient respiration. Hence, the invention makes use of the relationship between respiration and the level of C02 in the expired air. (Danehom i-f 11.) 3. Danehom teaches that "the level of C02 is measured over several breaths, and the respiratory frequency is extracted from the respiratory signal .... The respiratory frequency is then used to derive a model for the respiratory artefacts, and the model is subtracted from the measured blood pressure data" (Danehom i-f 13). 4. Danehom discloses a device for acquiring and correcting blood pressure data, having a catheter insertable into the heart or an artery for measuring blood pressure, the device further containing a sensor for measuring the level of C02 in the patient's expired air as a respiratory signal, and a data processing module for removing artefacts in the measured blood pressure data, wherein the data processing module is adapted to use the respiratory signal to 4 Appeal2014-000611 Application 12/638,658 approximate and to remove the artefacts caused by patient respiration. (Danehom i-f 20; see Ans. 4.) 5. Liao teaches the use of "a pulmonary artery pressure signal to detect and/or monitor physiological parameters" (Liao, Abstract). "[T]he invention includes a method of measuring a pulmonary function parameter of a patient including chronically implanting a pulmonary artery pressure sensor, obtaining a pulmonary artery pressure signal from the pressure sensor, and processing the signal to obtain the pulmonary function parameter" (Liao i-f 5; see Ans. 4). 6. Liao teaches that the "hemodynamic information that can be obtained with a coronary artery pressure sensor can include the systolic pulmonary artery pressure at end-expiration, the diastolic pulmonary artery pressure at end-expiration, [and] the mean pulmonary artery pressure," among others (Liao i-f 34). 7. Liao teaches that it has been discovered that a pulmonary artery pressure signal can be processed in order to determine or estimate one or more parameters of pulmonary function ("pulmonary function parameters"). Pulmonary artery pressure is modulated by intrathoracic pressure, which changes with inspiration and expiration. Specifically, intrathoracic pressure is increased during expiration and decreased during inspiration. The relationship between pulmonary artery pressure and intrathoracic pressure can be used in a method in order to derive one or more pulmonary function parameters. (Liao i-f 35; see Ans. 4.) 5 Appeal2014-000611 Application 12/638,658 8. Liao teaches that "[m]any different pulmonary function parameters can be calculated or estimated by processing a pulmonary artery pressure signal," including, for example, respiration waveforms, tidal volume, and forced expiration volume in one minute (FEV 1) (Liao i-f 3 7; see Ans. 4). 9. Figure 3 of Liao, reproduced below shows "an idealized pulmonary artery pressure signal and a respiration signal derived from the pulmonary artery pressure signal" (Liao i-f 13; see Ans. 5). "Respiration line 104 illustrates a roughly sinusoidal respiratory artifact that is superposed on pulmonary artery pressure and is caused by changes in intrathoracic pressure during the respiration cycle" (id. i-f 39). "The contours of the respiration line 104 over time can be referred to as the respiration waveform. The slope of the respiration line 104 as it is rising (as during expiration) and as it is falling (as during inspiration) can be tracked and recorded" (id. i-f 40). "The time for each cycle of respiration (both expiration and inspiration) can be determined by measuring the amount of time 106 in 6 Appeal2014-000611 Application 12/638,658 between successive peaks (or valleys) of the respiration line 104" (id. i-f 41; see Ans. 4). 10. Liao explains that "[ d]ata from the air flow meter can be used to accurately calculate the actual tidal volume. The recorded pulmonary artery signal can then be calibrated to the actual tidal volume as indicated by the air flow meter" (id. i-f 43). Analysis The Examiner finds that Danehom discloses "a device with two sensors; a catheter for sensing invasive blood pressure, and a sensor for measuring C02 as a respiratory signal" (Ans. 3--4 (citing Danehom, Abstract and i-f 2)). The Examiner finds that Danehom discloses "a data processing module for removing respiratory artifacts, in the invasive blood pressure measurements, using the respiratory data" (Ans. 4 (citing Danehom, Abstract, i-fi-12 and 20); see FF 1--4). According to the Examiner, "Danehom's device involves identification of respiratory frequency ... and/or heart rate" (Ans. 4 (citing Danehom, Abstract, i-fi-128, 29, 46, and 4 7) ). The Examiner finds that Danehom' s device determines "the actual blood pressure by removing respiratory disturbances from the measured blood pressure" (Ans. 4 (citing Danehom i-fi-130-32, Fig. 3, and Claim 16); see FF 4). The Examiner finds that "Danehom does not teach using the removed respiratory disturbances as an obtained respiratory function signal" (Ans. 4) (emphasis added). In other words, Danehom does measure respiratory frequency, but the reference does not keep the data from each sensor separately and uses the respiratory frequency to remove artifacts from the blood pressure signal (FF 4). The Examiner finds that Liao discloses "a 7 Appeal2014-000611 Application 12/638,658 method of measuring a pulmonary function parameter by processing a pulmonary artery pressure signal" (Ans. 4 (citing Liao i-fi-15, 35, and 37 and Claim 1 ); see FF 5-9). According to the Examiner, "[ w ]hat Liao does add to Danehom, is keeping the respiratory data extracted from measured invasive blood pressure data as a respiration signal, rather than removing it as noise" (Ans. 9-10 (citing Liao i-fi-15 and 35 and Claim 1)). Appellant argues that Danehom discloses "a device for obtaining blood pressure data ... [and] the present application is directed to an apparatus that measures a respiratory function. Thus, Danehom is direct[ ed] to a different type of apparatus" than the instantly claimed invention (App. Br. 11 ). Appellant argues that Danehom does not "teach or suggest 'a second sensor configured to measure frequency of at least one of a heart beat and a respiration,' as recited in claim 1" (App. Br. 11-12; see also Reply Br. 3). Appellant argues that because "Danehom does not provide a 'frequency measured by the second sensor,"' Danehom does not "teach or suggest 'a controller configured to obtain a respiratory function signal from the invasive blood pressure detected by the first sensor, by using at least one of the frequency measured by the second sensor and a harmonic of the frequency,' as recited in claim 1" (App. Br. 12; see also Reply Br. 3). Appellant argues that Liao "does not disclose 'a second sensor configured to measure frequency of a least one of a heart beat and a respiration"' (App. Br. 12; see also Reply Br. 3). Appellant's arguments are not persuasive. "The combination of familiar elements according to known methods is likelv to be obvious when ~ ~ it does no more than yield predictable results.~' KSR Int 'l Co. v. Tele/lex 8 Appeal2014-000611 Application 12/638,658 inc., 550 U.S. 398, 416 (2007). Liao discloses that the measurement of pulmonary artery blood pressure can be used to obtain both hemodynamic information and information on pulmonary function parameters due to the fact that pulmonary artery pressure is modulated by intrathoracic pressure, which changes with inspiration and expiration (FFs 5-10). Liao discloses that the relationship between pulmonary artery pressure and intrathoracic pressure can be used to derive one or more pulmonary function parameters (FF 7). Although Liao does not disclose the use of two separate sensors, Danehom discloses a device for removing artefacts caused by patient respiration in blood pressure measurement data that uses one sensor to measure blood pressure in an artery and a second sensor to measure expiration C02 and provide the respiration frequency (FFs 1--4). Because Liao and Danehom both disclose analyzing a blood pressure signal to determine the respiratory component of a blood pressure measurement, it would have been obvious to one of skill in the art to use Danehom's apparatus in Liao' s method so that respiration frequency as determined by C02 measurement could be used to verify the respiration component that is observed in Liao's analysis of pulmonary artery pressure (FF 9). Appellant's arguments that Danehom does not disclose a second sensor for measuring respiration frequency is also not persuasive because Danehom discloses that the C02 level is measured over several breaths to provide a respiratory frequency, which is then used to model and remove the respiratory artefacts in the blood pressure measurement (FF 3). 9 Appeal2014-000611 Application 12/638,658 Thus, we affirm the rejection of claim 1 under 35 U.S.C. § 103(a). Claims 2-10 have not been argued separately and therefore fall with claim 1. See 37 C.F.R. § 41.37(c)(l)(iv). II. Issue The Examiner has rejected claim 11 under 35 U.S.C. § 103(a) as obvious over Danehom, Liao, and Joeken (Ans. 6-7). The issue is: Does the evidence of record support the Examiner's conclusion that the combination ofDanehom, Liao, and Joeken would have made obvious a respiratory function measuring apparatus comprising "a respiration determiner configured to ... determine whether the respiration is spontaneous respiration or artificial respiration," as claimed? Findings of Fact 11. Joeken discloses an apparatus that "determines the breathing cycle from the central venous pressure signal. ... Summation/integration over (at least part of) the respiratory power spectrum delivers the respiratory power" (Joeken i-f 48). Joeken discloses that the "heart rate is determined from the arterial pressure signal . . . [and the] cardiac power spectrum is determined" (Joeken i-f 49). "[T]he ratio of respiratory and cardiac power is provided as a measure of volume responsiveness" (Joeken i-f 50). 12. Joeken discloses that the "power spectra can be characterized with respect to the kind of breathing (mechanically ventilated or spontaneous)" (Joeken i-f 52). "Controlled ventilation typically differs from spontaneous breathing by a smaller full width at half maximum WR and a 10 Appeal2014-000611 Application 12/638,658 slower decrease of log SR(t) with respect of log fat higher frequencies" (Joeken i-f 53). Analysis Appellant argues that Joeken discloses a "blood pressure measurement device and, as such, fails to make up for the ... deficiencies of Danehom and Liao" (App. Br. 13). Appellant argues that Joeken does not "disclose or teach at least the feature of 'a controller configured to obtain a respiratory function signal from the invasive blood pressure detected by the first sensor' as recited in claim 1" (id.). Appellant's argument is not persuasive because, as discussed above, we have concluded that claim 1 would have been obvious in view of the Danehom and Liao. Appellant argues that "it would not have been obvious to combine and modify Danehom, Liao and Joeken to arrive at the subject matter of claim 11" (App. Br. 13). Appellant argues that the rejection is conclusory and relies on impermissible hindsight (id.). The Examiner responds that all three references, Danehom, Liao, and Joeken, disclose "extracting respiration data flro]m invasive blood pressure [measurements], making them analogous art" (Ans. 10). According to the Examiner, "Joeken adds to Danehom and Liao by teaching identification of spontaneous vs. artificial respiration as an improvement on such extraction methods due to known differences between spontaneous and artificial respiration" (Ans. 10 (citing Joeken, Abstract, i-fi-148-54, Figs. 3 and 4); see FFs 11 and 12). 11 Appeal2014-000611 Application 12/638,658 We agree with the Examiner's reasoning and conclusion. As discussed above, Liao discloses monitoring of pulmonary artery pressure to estimate parameters of pulmonary function. J oeken discloses monitoring central venous pressure to determine the respiratory power spectrum (FF 11 ), and discloses that the respiratory power spectrum will indicate whether breathing is spontaneous or mechanically ventilated (FF 12). We agree with the Examiner that it would have been obvious to modify the apparatus made obvious by Danehom and Liao to further comprise a respiration determiner configured to determine whether the respiration is spontaneous or artificial because Joeken discloses that such information can be obtained by monitoring blood pressure, and one of skill the art would have appreciated that determining the mode of breathing, i.e., either spontaneous or artificial to be relevant to the determination of pulmonary function parameters (see Ans. 7). Accordingly, we affirm the rejection of claim 11 under 35 U.S.C. § 103(a) over the combination of Danehom, Liao, and Joeken. III. Issue The Examiner has rejected claims 12 and 15 under 35 U.S.C. § 103(a) as obvious in view of Danehom, Liao, and Euliano (Ans. 7-8). The issue is: Does the evidence of record support the Examiner's conclusion that the combination of Danehom, Liao, and Euliano would have made obvious the respiratory function measuring apparatus as claimed? 12 Appeal2014-000611 Application 12/638,658 Findings of Fact 13. Euliano discloses a "method of estimating the actual patient effort parameter using a combination of multiple parameters derived from sensors that monitor the patient and/or ventilator" (Euliano i-f 19). "The patient effort parameter can be any parameter that represents the effort exerted by the patient to breathe, including but not limited to work of breathing, power of breathing, or pressure time product" (Euliano i-f 19). 14. Euliano discloses that the parameters are preferably derived from the airway pressure, flow, and volume waveforms and the carbon dioxide and pulse oximeter waveforms normally collected by a respiratory monitor, including but not limited to tidal volume, breathing frequency, peak inspiratory pressure (PIP), inspiratory time, PO .1, trigger time, trigger depth, respiratory system resistance, respiratory compliance, end-tidal carbon dioxide, variations in the pulse oximeter plethysmogram, and the concavity/convexity of the pressure waveform. (Euliano i-f 20.) 15. Euliano discloses that, "[i]n the embodiment depicted in FIG. 2, a patient 10 requiring respiratory support and connected to a ventilator 12 will have an airway flow and pressure sensor 14, along with possibly a carbon dioxide detector attached at the y-piece of the standard ventilator circuit 16" (Euliano i-f 40). "These sensors measure the flow, pressure, and partial pressure of carbon dioxide in the gases that pass to and from the patient. ... These signals are then processed in a parameter extraction module 22 to calculate a variety of other parameters from the flow, pressure, and C02 data" (Euliano i-f 40). "For example ... peak inspiratory pressure is calculated by determining the maximum pressure during a breath; PO. I is 13 Appeal2014-000611 Application 12/638,658 calculated by measuring the change in airway pressure during the first tenth of a second of a breath" (Euliano i-f 40). Analysis Appellant argues that "it would not have been obvious to combine and modify Danehom, Liao and Euliano to arrive at the subject matter of claims 12 and 15 because they are devices directed to completely different end result objectives, i.e., blood pressure measurement and work of breathing measurement" (App. Br. 14). Appellant argues that one could not arrive at the present claims without the use of impermeable hindsight (id.). Appellant's arguments are not persuasive. As discussed above (J.), Liao discloses monitoring of pulmonary artery pressure to estimate parameters of pulmonary function. Danehom in combination with Liao would have made obvious the use of a second sensor to monitor expiration C02 so that respiration frequency, as determined by C02 measurement, could be used to verify the respiration component that is observed in Liao' s analysis of pulmonary artery pressure. Euliano discloses that a respiratory flow and pressure sensor combined with a C02 sensor can be used to provide additional pulmonary parameters such as peak inspiratory pressure and PO. I, which is the change in airway pressure during the first tenth of a second of a breath (FF 13-15). We agree with the Examiner's conclusion that it would have been obvious to include flow and pressure sensors to the apparatus of Liao and Danehom in order to monitor additional parameters for improved detection of irregular or interrupted breathing in a ventilated patient (Ans. 8). Accordingly, we affirm the rejection of claims 12 and 15 under 35 U.S.C. § 103(a) over the combination ofDanehom, Liao, and Euliano. 14 Appeal2014-000611 Application 12/638,658 Issue IV. The Examiner has rejected claims 13 and 14 under 35 U.S.C. § 103(a) as obvious in view ofDanehom, Liao, and Anderson (Ans. 8-11). The issue is: Does the evidence of record support the Examiner's conclusion that the combination ofDanehom, Liao, and Anderson would have made obvious a respiratory function measuring apparatus comprising a display to show the end-tidal? Findings of Fact 16. The Specification discloses that the end-tidal is the inspiration starting point (Spec. 13). 17. Liao discloses that graphs of respiration signals can be used to show respiratory parameters (Liao, Figs. 4, 5, 7-11, and 13-15). 18. Figure 4 of Liao is shown below: FIG. 4 is "a graph showing an idealized respiration signal during normal breathing and during a forced expiration maneuver" (Liao i-f 14). Liao discloses that, in Figure 4, "a graph is shown of a respiration signal 150 during normal breathing 152 and during forced expiration 154. The forced expiration amplitude 158 of the respiration signal 150 can be tracked and 15 Appeal2014-000611 Application 12/638,658 compared with the normal breathing amplitude 156 of the respiration signal 150" (Liao ii 46). 19. Anderson discloses a "method of graphically displaying cardiopulmonary function data by providing a plurality of axes radially projecting from a common point of origin" (Anderson, Abstract). "The cardiopulmonary function data to be displayed are then plotted ... [and] deviations of any such data from normal will result in an irregular polygon ... [which] provides a clinician with a readily visible display of possible cardiopulmonary d[y]sfunction" (Anderson, Abstract). 20. Anderson discloses that, although tabular data and plots can be interpreted with the trained eye, "that task is simplified by a display method which tends to magnify or highlight deviations of actual measured parameters from the norm" (Anderson, col. 3, 11. 55---60). Analysis Appellant contends that "Anderson also fails to supply the deficiencies of Danehom and Liao as described above in connection with claim 1" (App. Br. 15). Appellant argues that "it would not have been obvious to combine and modify Danehom and Liao with Anderson to arrive at the subject matter of claims 13 and 14" (id.). Appellant's arguments are not persuasive. Liao discloses the use of graphs of the respiratory signal or waveform to display and highlight functional parameters of respiration (FFs 17-18). One of skill the art would appreciate that marks highlighting the end-tidal volume would be useful to indicate changes to the respiratory signal. Anderson discloses that data interpretation can be simplified by display methods using graphs that 16 Appeal2014-000611 Application 12/638,658 magnify and highlight particular data features (FFs 19-20). In view of this disclosure in Liao and Anderson, we agree with the Examiner's conclusion that the combination of Danehom, Liao, and Anderson would have made obvious a respiratory function apparatus with displays configured to show the inspiration stating point. Accordingly, we affirm the rejection of claims 13 and 14 under 35 U.S.C. § 103(a) over the combination ofDanehom, Liao, and Anderson. SUMMARY We affirm the rejection of claims 1-15 under 35 U.S.C. § 103(a). No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). AFFIRMED 17 Copy with citationCopy as parenthetical citation